To determine the concentrations of nine drugs used in the treatment of cardiovascular diseases in human plasma through QuEChERS pre-treatment combined with ultra-performance liquid chromatography-tandem mass spectrometry. Plasma samples were extracted with 3 mL of acetonitrile, 400 mg of anhydrous magnesium sulfate as a salting agent, and 20 mg of C18 as a sorbent. An Agilent Poroshell 120 EC-C18 column (4.6 × 100 mm, 2.7 μm) was selected and methanol-0.1 % water was used as the mobile phase, and ESI positive ion detection mode was selected. Results: The plasma concentrations of nisoldipine, metoprolol, and prazosin exhibited good linearity within the range of 0.05–4.0 ng/mL (r > 0.99), while atenolol, bisoprolol, propranolol, rosuvastatin, and atorvastatin showed linearity within the range of 0.5–40 ng/mL (r > 0.99). Fluvastatin showed good linearity within the range of 5.0–400 ng/mL. The accuracy of the method ranged from 94.15 to 110.62 %, while the recovery levels were in the range of 85.23 %–115.13 %. The matrix effects were observed between-6.54 % and 12.43 %. The intra-day and inter-day RSD was <15 % for the three concentrations of low, medium, and high. Conclusion The proposed method is rapid, accurate, specific, simple, reproducible, and suitable for the simultaneous measurement of the concentrations of nine drugs used in the treatment of cardiovascular diseases in human plasma.
Here, a QuEChERS (quick, easy, cheap, effective, rugged, and safe) pretreatment method was combined with UPLC-MS/MS to facilitate the rapid and reliable simultaneous detection of five calcium channel blockers (CCBs) in human plasma. For this approach, samples were treated with 1 mL of acetonitrile, 350 mg of magnesium sulfate, and 70 mg of PSA adsorbent prior to centrifugation. Supernatants then underwent gradient elution for 8 min with an Agilent C18 column using an acetonitrile-water solution supplemented with 5 mmol⋅L−1 of ammonium acetate. This technique exhibited a good linear response in the 1–800 ng⋅mL−1 range for the analyzed drugs, with an R2≥ 0.9921, an accuracy of 87.54–113.05%, a matrix effect (ME) of 91.21–116.39%, a precision of 0.19–11.64%, and stability of no more than 10.05%. This time-saving QuEChERS reagent-based pretreatment technique thus allowed for the simultaneous and accurate detection of five CCBs in human plasma samples, providing a promising new basis for therapeutic drug monitoring in patients with hypertension.