Two optical waveguide sensors based on SOS (silicon-on-sapphire) for detecting CO2 are theoretically proposed. The operational wavelength is 4.23 μm, which is the maximum absorption line of CO2. The power confinement factor (η) value is over 40% and 50%, the propagation loss is 0.98 dB/cm and 2.99 dB/cm, respectively, in the slot waveguide and SWGS (subwavelength grating slot) waveguide. An inverted tapered structure is used for the transition from strip waveguide to slot waveguide and constitutes the sensing absorption region, with the coupling efficiency that can reach more than 90%. When the optimal absorption length of the slot waveguide and SWGS waveguide is 1.02 cm and 0.33 cm, respectively, the maximum sensitivity can reach 6.66 × 10−5 (ppm−1) and 2.60 × 10−5 (ppm−1). Furthermore, taking the slot waveguide as an example, spiral and meander structures enable the long-distance sensing path to integrate into a small area.
In the field of clinical medicine, the real-time monitoring of carbon dioxide (CO2) gas exhaled by the human body is of great significance. At present, the detection devices on the market mainly work by sucking a small amount of gas from the nasal cavity into the detection device, and they present some problems such as overly long sampling tubes, easy blockage or distortion, and an abnormal gas dispersion. In this paper, a micro/nano optical fiber sensor that can directly detect respiration is proposed and is characterized by the relative change in CO2 concentration. The measurement is achieved by using the principle of high evanescent field absorption, and the operating band is 2.004 μm. The sensing effects of micro/nano fibers with different diameters are compared, and the experiment proves that a micro/nano fiber with a diameter of 1 μm can be measured. The light intensity change over 30 s was measured, and the breathing rate was 10–11 times/min. The sensor can achieve a rapid real-time response to CO2 gas detection and is small in size, low-cost, and easy to replace. It has a great application potential in clinical scenarios such as in gastrointestinal endoscopes that require real-time monitoring of human respiration.
In the field of clinical medicine, the real-time monitoring of carbon dioxide gas exhaled by the human body is of great significance. At present, the detection devices in the market are mainly detected by sucking a small amount of gas in the nasal cavity to the detection device, and there are some problems such as too long sampling tubes, easy blockage or distortion, and abnormal gas dispersion. In this paper, a micro-nano optical fiber sensor that can directly detect the concentration of end-tidal carbon dioxide is proposed. The measurement is achieved by using the principle of high evanescent field absorption, and the operating band is 2.004 μm. The sensor uses micro-nano optical fiber as the sensing area, and then detects the presence of carbon dioxide gas exhaled by human body through optical power attenuation. The function of micro-nano fiber is to realize the transmission of signal light and also serve as the absorption medium of the gas to be measured. In addition, the variation of light power also reflects the respiratory cycle of the human body. The sensor can realize rapid real-time response to carbon dioxide gas detection, with small size, low cost, and easy to replace. It has great application potential in clinical scenarios such as Gastrointestinal Endoscopes that require real-time monitoring of human respiration.