A modified two-axis surface encoder is proposed to separately measure both the in-plane displacement and the Z-directional out-of-plane displacement with minor crosstalk errors. The surface encoder is composed of a scale grating and a small-sized sensor head. In the modified surface encoder, the measurement laser beam from the sensor head is designed to be projected onto the scale grating at a right angle. For measurement of the X- and Y-directional in-plane scale displacement, the positive and negative first-order diffracted beams from the scale grating are superimposed on each other in the sensor head, producing interference signals. On the other hand, the Z-directional out-of-plane scale displacement is measured based on the principle of a Michelson-type interferometer. To avoid the influence of reflection from the middle area of the transparent grating, which causes periodic crosstalk errors in the previous research, a specially fabricated transparent grating with a hole in the middle is employed in the newly designed optical system. A prototype sensor head is constructed, and basic performances of the modified surface encoder are tested by experiments.
This paper proposed angle measurement methods based on direct third harmonic generation (THG) in centrosymmetric crystals. The principles of the intensity-dependent and the wavelength-dependent angle measurement methods were illustrated. In this study, three prospective centrosymmetric crystals and two different phase-matching types were investigated in a wavelength range from 900 nm to 2500 nm. For the intensity-dependent method, a dispersion-less wavelength range was found from 1700 nm to 2000 nm for α-BBO and calcite. Compared with rutile, α-BBO and calcite had relatively better measurement performance based on the angle measurement sensitivity calculation. The wavelength-dependent method was considered in a dispersive range of around 1560 nm. The results suggested that α-BBO and calcite were also suitable for wavelength-dependent measurement. In addition, the effects of focusing parameters were considered in the simulation, and the optimized focal length (f = 100 mm) and the focused position (in the center of the crystal) were determined.