Yue Li

Mrs. Yue Li


Research Focus & Metrics

Academic Career Length

4Years (2023-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
5 September 2023 in Photonics
Verified
Journal Article
Open access
PDF file
YLYue Li
Yue Li
ZLZhen Liu
Zhen Liu
YYYuan Yinlin
Yuan Yinlin
WZWenchao Zhai
Wenchao Zhai
PZPeng Zou
Peng Zou
XZXiaobing Zheng
Xiaobing Zheng
In order to establish a transferring chain from a photon flux of a single-photon source in quantum radiometry, the nonlinearity of the photodetector needs to be accurately measured. Using the flux superposition method, a nonlinearity measurement setup has been designed. The measurement setup consists of two tungsten halogen lamps, parent–child integrating spheres, an adjustable aperture, a diaphragm tube, and an optical filter. It has the advantage of low polarization error, low interference error, and low stray light effect. The Si photodiode to be measured is cooled to −40 °C to obtain a low noise level for low-flux radiation measurement. The nonlinearity of the Si photodetector is measured for photocurrent ranges from 10−12 A~10−6 A level, with a relative standard uncertainty from 0.0092~0.023%. The relative standard uncertainty of the nonlinearity correction factor ranged from 0.023~0.049%.