19 April 2022 in Micromachines
Verified
Journal Article
Open access
PDF file
POProsper Obaro Oyibo
Prosper Obaro Oyibo
SJSatyajith Jujjavarapu
Satyajith Jujjavarapu
BMBrice Meulah
Brice Meulah
TATope Ebenzer Agbana
Tope Ebenzer Agbana
IBIngeborg Braakman
Ingeborg Braakman
JDJan Carel Diehl
Jan Carel Diehl
For many parasitic diseases, the microscopic examination of clinical samples such as urine and stool still serves as the diagnostic reference standard, primarily because microscopes are accessible and cost-effective. However, conventional microscopy is laborious, requires highly skilled personnel, and is highly subjective. Requirements for skilled operators, coupled with the cost and maintenance needs of the microscopes, which is hardly done in endemic countries, presents grossly limited access to the diagnosis of parasitic diseases in resource-limited settings. The urgent requirement for the management of tropical diseases such as schistosomiasis, which is now focused on elimination, has underscored the critical need for the creation of access to easy-to-use diagnosis for case detection, community mapping, and surveillance. In this paper, we present a low-cost automated digital microscope—the Schistoscope—which is capable of automatic focusing and scanning regions of interest in prepared microscope slides, and automatic detection of Schistosoma haematobium eggs in captured images. The device was developed using widely accessible distributed manufacturing methods and off-the-shelf components to enable local manufacturability and ease of maintenance. For proof of principle, we created a Schistosoma haematobium egg dataset of over 5000 images captured from spiked and clinical urine samples from field settings and demonstrated the automatic detection of Schistosoma haematobium eggs using a trained deep neural network model. The experiments and results presented in this paper collectively illustrate the robustness, stability, and optical performance of the device, making it suitable for use in the monitoring and evaluation of schistosomiasis control programs in endemic settings.

4 October 2018 in PLOS ONE
Verified
Journal Article
Open access
PDF file
TATope Ebenzer Agbana
Tope Ebenzer Agbana
JDJan Carel Diehl
Jan Carel Diehl
FPFiona van Pul
Fiona van Pul
SKShahid M Khan
Shahid M Khan
VPVsevolod Patlan
Vsevolod Patlan
GVGleb Vdovin
Gleb Vdovin
We have optimized the design and imaging procedures, to clearly resolve the malaria parasite in Giemsa-stained thin blood smears, using simple low-cost cellphone-based microscopy with oil immersion. The microscope uses a glass ball as the objective and the phone camera as the tube lens. Our optimization includes the optimal choice of the ball lens diameter, the size and the position of the aperture diaphragm, and proper application of immersion, to achieve diagnostic capacity in a wide field of view. The resulting system is potentially applicable to low-cost in-the-field optical diagnostics of malaria as it clearly resolves micron-sized features and allows for analysis of parasite morphology in the field of 50 × 50 μm, and parasite detection in the field of at least 150 × 150 μm.

6 June 2017 in Optics Letters
Verified
Journal Article
Closed access
PDF file
TATope Ebenzer Agbana
Tope Ebenzer Agbana
HGHai Gong
Hai Gong
AAAbena S. Amoah
Abena S. Amoah
VBVitaly Bezzubik
Vitaly Bezzubik
MVMichel Verhaegen
Michel Verhaegen
GVGleb Vdovin
Gleb Vdovin
We have shown that the maximum achievable resolution of an in-line lensless holographic microscope is limited by aliasing and, for collimated illumination, cannot exceed the camera pixel size. This limit can be achieved only when the optimal conditions on the spatial and temporal coherence state of the illumination are satisfied. The expressions defining the configuration, delivering maximum resolution with given spatial and temporal coherence of the illumination, are obtained. The validity of these conditions is confirmed experimentally.

17 October 2022 in The American Journal of Tropical Medicine and Hygiene
Verified
Journal Article
Open access
PDF file
BMBrice Meulah
Brice Meulah
POProsper Obaro Oyibo
Prosper Obaro Oyibo
MBMichel Bengtson
Michel Bengtson
TATope Ebenzer Agbana
Tope Ebenzer Agbana
RLRoméo Aimé Laclong Lontchi
Roméo Aimé Laclong Lontchi
LVLisette Van
Lisette Van
: Conventional microscopy is the standard procedure for the diagnosis of schistosomiasis, despite its limited sensitivity, reliance on skilled personnel, and the fact that it is error prone. Here, we report the performance of the innovative (semi-)automated Schistoscope 5.0 for optical digital detection and quantification of Schistosoma haematobium eggs in urine, using conventional microscopy as the reference standard. At baseline, 487 participants in a rural setting in Nigeria were assessed, of which 166 (34.1%) tested S. haematobium positive by conventional microscopy. Captured images from the Schistoscope 5.0 were analyzed manually (semiautomation) and by an artificial intelligence (AI) algorithm (full automation). Semi- and fully automated digital microscopy showed comparable sensitivities of 80.1% (95% confidence interval [CI]: 73.2–86.0) and 87.3% (95% CI: 81.3–92.0), but a significant difference in specificity of 95.3% (95% CI: 92.4–97.4) and 48.9% (95% CI: 43.3–55.0), respectively. Overall, estimated egg counts of semi- and fully automated digital microscopy correlated significantly with the egg counts of conventional microscopy (r = 0.90 and r = 0.80, respectively, P < 0.001), although the fully automated procedure generally underestimated the higher egg counts. In 38 egg positive cases, an additional urine sample was examined 10 days after praziquantel treatment, showing a similar cure rate and egg reduction rate when comparing conventional microscopy with semiautomated digital microscopy. In this first extensive field evaluation, we found the semiautomated Schistoscope 5.0 to be a promising tool for the detection and monitoring of S. haematobium infection, although further improvement of the AI algorithm for full automation is required.

8 September 2023 in Scientific Reports
Verified
Journal Article
Open access
PDF file
LMLouise Makau-Barasa
Louise Makau-Barasa
LALiya Assefa
Liya Assefa
MAMoses Oluwaseun Aderogba
Moses Oluwaseun Aderogba
DBDavid Bell
David Bell
JSJacob Solomon
Jacob Solomon
TATope Ebenzer Agbana
Tope Ebenzer Agbana
In this research, we report on the performance of automated optical digital detection and quantification of Schistosoma haematobium provided by AiDx NTDx multi-diagnostic Assist microscope. Our study was community-based, and a convenient sampling method was used in 17 communities in Abuja Nigeria, based on the disease prevalence information extracted from the baseline database on schistosomiasis, NTD Division, of the Federal Ministry of Health. At baseline, samples from 869 participants were evaluated of which 358 (34.1%) tested S. haematobium positive by the reference diagnostic standard. Registered images from the fully automated (autofocusing, scanning, image registration and processing, AI image analysis and automatic parasite count) AiDx assist microscope were analyzed. The Semi automated (autofocusing, scanning, image registration & processing and manual parasite count) and the fully automated AiDx Assist showed comparable sensitivities and specificities of [90.3%, 98%] and [89%, 99%] respectively. Overall, estimated egg counts of the semi-automated & fully automated AiDx Assist correlated significantly with the egg counts of conventional microscopy (r = 0.93, p ≤ 0.001 and r = 0.89, p ≤ 0.001 respectively). The AiDx Assist device performance is consistent with requirement of the World Health Organization diagnostic target product profile for monitoring, evaluation, and surveillance of Schistosomiasis elimination Programs.

1 July 2023 in Journal of Medical Imaging
Verified
Journal Article
Open access
PDF file
POProsper Obaro Oyibo
Prosper Obaro Oyibo
BMBrice Meulah
Brice Meulah
MBMichel Bengtson
Michel Bengtson
LVLisette Van
Lisette Van
WOWellington Aghoghovwia Oyibo
Wellington Aghoghovwia Oyibo
TATope Ebenzer Agbana
Tope Ebenzer Agbana
Schistosomiasis is endemic in 76 countries worldwide with ∼252  million people infected and an estimated 779 million people at risk of infection.1 Schistosomiasis is caused by blood flukes of the genus Schistosoma (S); both S. mansoni (intestinal schistosomiasis) and S. haematobium (urogenital schistosomiasis) are endemic in Africa.2 Schistosomiasis presents a substantial public health and economic burden as it is a disease of poverty. In the drive to attain the World Health Organization (WHO) control and elimination targets, diagnosis for adequate monitoring of interventions and surveillance is critical.2,3 Recently, the WHO published the diagnostic target product profiles (TPP) for monitoring, evaluation, and surveillance of schistosomiasis control programs,4 which identifies development of diagnostic tests for S. haematobium detection as a high-risk requirement due to lack of its availability. The TPP suggests a semi-quantitative analysis, capable of providing some degree of information regarding intensity of infection, as ideal for a diagnostic test for schistosomiasis to support monitoring and evaluation.4 Currently, microscopy is the WHO reference standard for the diagnosis of schistosomiasis in resource-limited settings. For the detection of S. haematobium infection, urine samples, after filtration, sedimentation, or centrifugation, are microscopically examined for the presence of eggs.3 This method is operator dependent, costly, laborious, and time-consuming. Furthermore, it requires expertise, which means microscopy skills need to be gained and maintained, which can be an economic challenge, particularly in remote rural communities.3 There is also the risk of visual health complications among microscopists resulting from excessive workload due to the low ratio of trained microscopists to samples for analysis in endemic regions.5 Hence, a field adaptable, rapid, and easy-to-use automated diagnosis is relevant for the prompt detection of cases, which will facilitate mapping and monitoring of interventions.4 Recent advances in opto-mechanics and opto-electronics have rapidly transformed the field of biomedical optics. Optical imaging technologies, such as conventional light microscopes, are being redesigned to integrate and miniaturize portable light microscopes for use at the point of care.6 – 9 Although these technologies are readily available in high-income countries, unfortunately, nearly all schistosomiasis cases are seen in low-resource regions of low-income countries, significantly justifying the need for cost-effective and easy-to-use smart diagnostic technologies. In this work, we address these challenges by first increasing the size of the S. haematobium (SH) dataset in our previous work10 from 5198 to 12,051 images of clinical samples.11 We carry out detection and counting of S. haematobium eggs present in each image by proposing a two-stage framework consisting of a DeepLabv3 with MobilenetV3 backbone deep convolutional neural network12 trained on the SH dataset using a transfer learning approach. The second stage of our proposed framework is a refined segmentation and egg counting procedure, which adapts the region-based fitting of overlapping ellipses13 to efficiently separate the boundaries of overlapping eggs in the image. Finally, the detected isolated eggs are screened for the presence of an egg, which meets the defined boundary condition before the sample can be determined as positive/negative diagnosis. We further demonstrate the robustness and applicability of the proposed framework for field diagnoses of urogenital schistosomiasis by implementing our framework on an Edge AI system (Raspberry Pi + Coral USB accelerator) and testing 65 clinical urine samples obtained in a field settings in Nigeria. The main contributions of this work can be summarized as follows.