Drift tube ion mobility spectrometers (IMS) are widely used for fast trace gas detection in air, but portable compact systems are typically very limited in their resolving power. Decreasing the initial ion packet width improves the resolution, but is generally associated with a reduced signal-to-noise-ratio (SNR) due to the lower number of ions injected into the drift region. In this paper, we present a refined theory of IMS operation which employs a combined approach for the analysis of the ion drift and the subsequent amplification to predict both the resolution and the SNR of the measured ion current peak. This theoretical analysis shows that the SNR is not a function of the initial ion packet width, meaning that compact drift tube IMS with both very high resolution and extremely low limits of detection can be designed. Based on these implications, an optimized combination of a compact drift tube with a length of just 10 cm and a transimpedance amplifier has been constructed with a resolution of 183 measured for the positive reactant ion peak (RIP+), which is sufficient to e.g. separate the RIP+ from the protonated acetone monomer, even though their drift times only differ by a factor of 1.007. Furthermore, the limits of detection (LODs) for acetone are 180 pptv within 1 s of averaging time and 580 pptv within only 100 ms.
There is increasing evidence that breath volatile organic compounds (VOC) have the potential to support the diagnosis and management of inflammatory diseases such as COPD. In this study we used a novel breath sampling device to search for COPD related VOCs. We included a large number of healthy controls and patients with mild to moderate COPD, recruited subjects at two different sites and carefully controlled for smoking. 222 subjects were recruited in Hannover and Marburg, and inhaled cleaned room air before exhaling into a stainless steel reservoir under exhalation flow control. Breath samples (2.5 l) were continuously drawn onto two Tenax(®) TA adsorption tubes and analyzed in Hannover using thermal desorption-gas chromatography-mass spectrometry (TD-GC-MS). Data of 134 identified VOCs from 190 subjects (52 healthy non-smokers, 52 COPD ex-smokers, 49 healthy smokers, 37 smokers with COPD) were included into the analysis. Active smokers could be clearly discriminated by higher values for combustion products and smoking related VOCs correlated with exhaled carbon monoxide (CO), indicating the validity of our data. Subjects from the study sites could be discriminated even after exclusion of cleaning related VOCs. Linear discriminant analysis correctly classified 89.4% of COPD patients in the non/ex-smoking group (cross validation (CV): 85.6%), and 82.6% of COPD patients in the actively smoking group (CV: 77.9%). We extensively characterized 134 breath VOCs and provide evidence for 14 COPD related VOCs of which 10 have not been reported before. Our results show that, for the utilization of breath VOCs for diagnosis and disease management of COPD, not only the known effects of smoking but also site specific differences need to be considered. We detected novel COPD related breath VOCs that now need to be tested in longitudinal studies for reproducibility, response to treatment and changes in disease severity.
Due to its high sensitivity, compact size and low cost ion mobility spectrometry (IMS) has the potential to become a point-of-care breath analyzer. Therefore, we developed a prototype of a compact, closed gas loop IMS with gas chromatographic (GC) pre-separation and high resolving power of R = 90. In this study, we evaluated the performance of this GC-IMS under clinical conditions in a COPD study to find correlations between VOCs (10 ppb v to 1 ppm v ) and COPD. Furthermore, in order to investigate possible correlations between ultra-low concentrated breath VOCs (0.1 ppt v to 1 ppb v ) and COPD, a modified mass spectrometer (MS) with atmospheric pressure chemical ionization (APCI) and GC pre-separation (GC-APCI-MS) was used. The GC-IMS has been used in 58 subjects (21 smokers with moderate COPD, 12 ex-smokers with COPD, 16 healthy smokers and 9 non-smokers). GC-APCI-MS data were available for 94 subjects (21 smokers with moderate COPD, 25 ex-smokers with COPD, 25 healthy smokers and 23 non-smokers). For 44 subjects, a comparison between GC-IMS and GC-APCI-MS data could be performed. Due to service intervals, subject availability and corrupt data, patient numbers were different for GC-APCI-MS and GC-IMS measurements. Using GC-IMS, three VOCs have been found showing a significant difference between healthy controls and patients with COPD. In the GC-APCI-MS data, we only observed one distinctive VOC, which has been identified as 2-pentanone. This proof-of-principle study shows the potential of our high-resolution GC-IMS in the clinical environment. Due to different linear dynamic response ranges, the data of GC-IMS and GC-APCI-MS were only comparable to a limited extent.