Introduction and Hypothesis: The aim of this study is to develop and validate a new integral parameter, the Biomechanical Integrity score (BI-score), for the characterization of the female pelvic floor.Methods: A total of 253 subjects with normal and pelvic organ prolapse (POP) conditions were included in the multi-site observational, case-control study; 125 subjects had normal pelvic floor conditions, and 128 subjects had POP stage II or higher. A Vaginal Tactile Imager (VTI) was used to acquire and automatically calculate 52 biomechanical parameters for eight VTI test procedures (probe insertion, elevation, rotation, Valsalva maneuver, voluntary muscle contractions in two planes, relaxation, and reflex contraction). Statistical methods were applied (t-test, correlation) to identify the VTI parameters sensitive to the pelvic conditions.Results: Twenty-six parameters were identified as statistically sensitive to POP development. They were subdivided into five groups to characterize (1) tissue elasticity, (2) pelvic support, (3) pelvic muscle contraction, (4) involuntary muscle relaxation, and (5) pelvic muscle mobility. Every parameter was transformed to its standard deviation units against the patient age similar to T-score for bone density. Linear combinations with specified weights led to the composition of five component parameters for groups (1)–(5) and the BI-score in standard deviation units. Thep-value for the BI-score hasp = 4.3 × 10−31for POP versus normal conditions. A reference BI-score curve against age for normal pelvic floor conditions was defined.Conclusions: Quantitative transformations of the pelvic tissues, support structures, and functions under diseased conditions may be studied with the BI-score in future research and practical applications.
This study aims to explore the applicability of an innovative cervical tactile ultrasound approach for predicting spontaneous preterm birth (sPTB). Eligible participants were women with low-risk singleton pregnancies in their second trimester, enrolled in this prospective observational study. A Cervix Monitor (CM) device was designed with a vaginal probe comprising four tactile sensors and a single ultrasound transducer operating at 5 MHz. The probe enabled the application of controllable pressure to the external cervical surface, facilitating the acquisition of stress-strain data from both anterior and posterior cervical sectors. These data were used to calculate cervical elasticity, as a stress-to-strain ratio. CM examination data were analyzed for 127 women at 24 - 28 gestational weeks. PTB was observed in 6.3% (8 out of 127) of the cases. The preterm group exhibited a lower average cervical stress-to-strain ratio (elasticity) of 0.70±0.26 kPa/mm compared to the term group's 1.63±0.65 kPa/mm, with a significant p-value of 1.1×10-4. Diagnostic accuracy for predicting spontaneous preterm birth based solely on cervical elasticity data was found to be 95.0% (95% CI, 88.5–100.0). These findings suggest that measuring cervical elasticity with the designed tactile ultrasound probe could potentially predict spontaneous preterm birth in a cost-effective manner.