A pelvic endoprosthesis is commonly used in orthopedic surgeries to reconstruct the pelvis after internal hemipelvectomy. This study presents the detailed design of a novel type I+II+III adjustable hemipelvic prosthesis based on the geometrical features of massive human pelvises. Finite element analysis is conducted to estimate the biomechanical performance of the newly designed adjustable hemipelvic prosthesis. Detailed numerical models of the natural and reconstructed pelvises including related soft tissues are developed. Hip contact forces during normal walking, which is one of the most frequent dynamic activities in daily living, are imposed on the pelvis. Results show that the peak stress observed in the reconstructed pelvis model is still within a low and elastic range below the yielding strength of the cortical bone and Ti6Al4V. No significant difference of the stress transferring route, displacement distributions and principal stress vectors is observed between the reconstructed and natural pelvises. The results indicate that the load transferring function of the partially resected pelvis is able to be reliably recovered by the adjustable hemipelvic prosthesis. The principal stress vectors in both pelvis models predict that bone absorption may not apparently occur in the long run. Long-term biomechanical performance of this newly designed prosthesis may be stability.
In this study, the biomechanical characteristic of a newly developed adjustable hemipelvic prosthesis under dynamic loading conditions was investigated using explicit finite element method. Both intact and reconstructed pelvis models, including pelvis, femur and soft tissues, were established referring to human anatomic data using a solid geometry of a human pelvic bone. Hip contact forces during ascending stairs and descending stairs were imposed on pelvic models. Results showed that maximum von Mises stresses in reconstructed pelvis were 421.85 MPa for prostheses and 109.12 MPa for cortical bone, which were still within a low and elastic range below the yielding strength of Ti-6Al-4V and cortical bone, respectively. Besides, no significant difference of load transferring paths along pelvic rings was observed between the reconstructed pelvis and natural pelvis models. And good agreement was found between the overall distribution of maximum principal stresses in trabecular bones of reconstructed pelvis and natural pelvis, while at limited stances, principal stresses in trabecular bone of reconstructed pelvis were slightly lower than natural pelvis. The results indicated that the load transferring function of pelvis could be restored by this adjustable hemipelvic prosthesis. Moreover, the prosthesis was predicted to have a reliable short- and long-term performance. However, due to the occurrence of slightly lower principal stresses at a few stances, a porous structure applied on the interface between the prosthesis and bone would be studied in future work to obtain better long-term stability.