The velocity map imaging (VMI) technique is used to acquire the momentum distribution of charged particles. Here, we introduce two additional operation modes for our recently built velocity map imaging (VMI) spectrometer: the spatial mapping mode that magnifies the image of zero energy ions with different scales and the high-resolution momentum mapping mode that acquires the electron momentum distribution at the kinetic energy of about 100 eV. In simulations, the ion image is magnified with a factor of up to 7.6, and a relative resolution of 0.15% at 150 eV electron kinetic energy is predicted. Switching between these two modes helps reduce the alignment error to below 0.2 mm. In the test using the above-threshold ionization (ATI) of argon (Ar), the Ar+ ion image is magnified by a factor of up to 6.7, and a relative resolution of 1.3% at 44.6 eV electron kinetic energy is achieved.
We report on a co-axial velocity map imaging (VMI) spectrometer in which the laser beam propagates along the spectrometer axis and the laser polarization plane is parallel to the detector. As a result, polarization-driven processes are recorded with much more detail compared to normal VMI. Simulation using SIMION predicts an exceptional relative energy resolution of 0.3% at kinetic energy ∼60 eV. Experimental validation via above-threshold ionization of xenon demonstrates two key properties: (1) a relative energy resolution of 5.5% at 11.5 eV of kinetic energy and (2) direct observation of attosecond angular streaking effects in photoelectron momentum distributions. These capabilities establish the spectrometer as a powerful tool for attosecond experiments.