A transfer radiometer (TR) applied to an on-orbit radiometric benchmark transfer chain has been developed, which can achieve the high-precision transformation of power and radiance responsivity and transmit the radiance responsivity traced to the cryogenic radiometer to remote sensors, such as an imaging spectrometer, so that the on-orbit remote sensors can achieve the high accuracy calibration of 10−3 magnitude. Radiance comparison experiments between the TR and the radiance standard of the National Institute of Metrology (NIM) were carried out to demonstrate the absolute accuracy of the TR radiance measurement. At 780.0 nm and 851.9 nm, the relative measurement uncertainties of the TR filter-free channel were 0.24% (k = 1). Additionally, the radiance measurement results of the TR were consistent with those of the NIM radiance meter, and the radiance measurement results’ relative differences between the TR and the NIM radiance meter were approximately 0.04% at 780.0 nm and 851.9 nm. The relative measurement uncertainties of TR 780.4 nm and 851.8 nm filter channels were 0.89% (k = 1) and 0.84% (k = 1), respectively. Additionally, the radiance measurement results of the TR 780.4 nm and 851.8 nm filter channels were consistent with the radiances of the integrating sphere source calibrated by the NIM at 780.4 nm and 851.8 nm; the relative differences between the radiances measured by the two TR filter channels and the radiances of the integrating sphere source itself were better than 0.56%. This proved that the TR could measure the monochromatic source radiance with a measurement uncertainty of 0.24% and measure the broadband source radiance with a measurement uncertainty better than 0.89%. The TR can be applied to the radiometric benchmark transfer chain to improve the measurement precision of on-orbit remote-sensing instruments.
In order to meet the high-accuracy calibration requirements of satellite remote sensing instruments, a transfer radiometer for an on-orbit radiometric benchmark transfer chain has been developed, which provides vital technical support for realizing the radiometric calibration uncertainty of the order of 10−3 for remote sensing instruments. The primary role of the transfer radiometer is to convert from the spectral power responsivity traceable to a cryogenic radiometer to the spectral radiance responsivity and transfer it to the imaging spectrometer. At a wavelength of 852.1 nm, the combined uncertainty of the radiance measurement comparison experiment between the transfer radiometer and a radiance meter is 0.43% (k = 1), and the relative deviation of the measurements between the transfer radiometer and the radiance meter is better than 0.36%, which is better than the combined uncertainty of the radiance measurement comparison experiment. This demonstrates that the transfer radiometer can achieve radiance measurements with a measurement uncertainty better than 0.3% (k = 1).
This research project envisions using a lunar observation platform to measure the full-wave (0.2~100 μm) and shortwave (0.2~4.3 μm) radiation of the Earth, achieving an accurate estimation of the overall radiation budget of the Earth. Based on the lunar platform, the system analyzes Earth’s radiation characteristics and geometric attributes, as well as the sampling properties of observation times. Informed by these analyses, an Earth-facing optical radiation measurement system tailored to these specifications is designed. The optical system adopts an off-axis three-mirror configuration with a secondary image plane, incorporating a field stop at the primary image plane to effectively suppress solar stray light, scattered lunar surface light, and background radiation from the instrument itself, ensuring the satisfactory signal-to-noise ratio, detection sensitivity, and observation duration of the instrument. At the same time, stringent requirements are imposed for the surface treatments of instrument components and temperature control accuracy to further ensure accuracy. Simulation analyses confirm that the design satisfies requirements, achieving a measurement accuracy of better than 1% across the entire optical system. This Moon-based Earth-radiation measurement system, with capabilities for Earth-pointing tracking, radiation energy detection, and stray-light suppression, furnishes a more comprehensive dataset, helping to advance our understanding of the mechanisms driving global climate change
We propose an observation system based on the Lagrange L1 point of the Earth–Moon system to observe solar spectral radiation reflected from the Earth, enabling continuous hyperspectral observation of the Earth’s hemisphere. The system can observe the solar spectral radiation reflected by the Moon, with its data applicable to on-orbit spectral radiation calibration. In this paper, the spectral irradiance at the entrance pupil of the Earth spectral radiation observation system (ESROS) is analyzed, and the optical design of the ESROS is introduced. An off-axis two-mirror telescope system, a coupling system of a microlens array and a fiber bundle, and an optical splitting system based on concave grating are used to achieve the full field of view hyperspectral splitting and miniaturization of the instrument. Finally, the stray radiation suppression of the instrument is introduced. The results show that the spectral resolution of the system is better than 5 nm in the 380–1000 nm band, and the spectral resolution is better than 10 nm in the 1000–1700 nm band. When observing the Earth, the signal-to-noise ratio is greater than 200. The external stray radiation suppression reaches the order of 10−9. The ESROS will provide crucial data support for researching global energy balance, climate change, and the spectral characteristics of exoplanets, facilitating planetary science and the exploration of extraterrestrial life.
High-precision absolute radiometric calibration is the cornerstone of space-based quantitative remote sensing. Monochromator-based radiance calibration is a key approach for space-borne radiometric benchmark transfer; however, the influence of the monochromator bandwidth on calibration accuracy urgently requires validation via an independent, SI-traceable benchmark. Taking a spectrometer as a case study, this paper establishes an equivalent theoretical calibration model. The model reveals that discrepancies in the equivalent spectral distributions of the calibration sources primarily contribute to their calibration deviations. Within the 775–855 nm range, cross-comparison experiments were conducted between the monochromator facility and a reference standard lamp-plaque. Experimental results demonstrate that under bandwidth configurations of 2 nm, 4 nm, and 6 nm, relative deviations in calibrated radiance responsivity stabilize at approximately 2.3%, exhibiting no significant dependence on bandwidth variations. Crucially, the consistency between the two methodologies is rigorously verified, as the radiance responsivity ratios uniformly fall within the combined expanded measurement uncertainty bounds. By demonstrating that bandwidth expansion within the specified 2–6 nm range does not degrade calibration accuracy, this study provides a critical empirical basis for safely increasing solar monochromator bandwidths to improve the signal-to-noise ratio (SNR) in next-generation benchmark missions like the Space-based Radiometric Measurement Benchmark Mission (LIBRA) program.