Christian von Mitzlaff

Dr. Christian von Mitzlaff

orcid

Research Focus & Metrics

Academic Career Length

7Years (2020-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
4 November 2020 in Cochlear Implants International
Verified
Journal Article
Closed access
CMChristian von Mitzlaff
Christian von Mitzlaff
ADAdrian Dalbert
Adrian Dalbert
SWSebastian Winklhofer
Sebastian Winklhofer
DVDorothe Veraguth
Dorothe Veraguth
AHAlexander Huber
Alexander Huber
CRChristof Roosli
Christof Roosli
Objective: To investigate impedance change of cochlear implant (CI) electrodes from CochlearTM in patients with straight and precurved electrodes, and to determine if impedance change of the basal electrodes are indicative for cochlear electrode migration. Methods: Retrospective case series at an academic tertiary referral center. Five hundred sixty patients (414 precurved, 146 straight) received a CI from CochlearTM between January 2010 and August 2018. Electrode impedance measured intraoperatively and during follow-up has been assessed. In case of impedance change of the 3 most basal electrodes a computer tomography (CT) scan has been performed. Cochlear position of the electrode array has been determined and compared to the one day postoperative electrode position (plain radiography). Results: Changes in electrode impedance in the basal electrodes occurred in 2.1% ears (n = 12). CT scan was available in 9 patients and revealed a migration of the electrode array in 6 patients. Straight electrode types showed a significantly higher migration rate than precurved electrode types (4.1% vs 0%). Conclusions: Electrode impedance is a useful tool to investigate electrode array integrity as well as to raise suspicion of possible electrode migration. Patients implanted with a straight electrode from CochlearTM array might benefit from lead wire fixation.

29 August 2024 in Journal of Clinical Medicine
Verified
Journal Article
Open access
PDF file
RBRahel Bertschinger
Rahel Bertschinger
CMChristian von Mitzlaff
Christian von Mitzlaff
MGMarlies Geys
Marlies Geys
AKAhmet Kunut
Ahmet Kunut
IDIvo Dobrev
Ivo Dobrev
ADAdrian Dalbert
Adrian Dalbert
Background: Cochlear implant (CI) electrode insertion can change the mechanical state of the ear whereby wideband tympanometry absorbance (WBTA) may serve as a sensitive tool to monitor these mechanical changes of the peripheral auditory pathway after CI surgery. In WBTA, the amount of acoustic energy reflected by the tympanic membrane is assessed over a wide frequency range from 226 Hz to 8000 Hz. The objective of this study was to monitor changes in WBTA in CI recipients before and after surgery. Methods: Following otoscopy, WBTA measurements were conducted twice in both ears of 38 standard CI recipients before and in the range of 4 to 15 weeks after CI implantation. Changes from pre- to postoperative absorbance patterns were compared for the implanted as well as the contralateral control ear for six different frequencies (500 Hz, 750 Hz, 1000 Hz, 2000 Hz, 3000 Hz, 4000 Hz). Furthermore, the influence of the time point of the measurement, surgical access, electrode type, sex and side of the implantation were assessed for the implanted and the control ear in a linear mixed model. Results: A significant decrease in WBTA could be observed in the implanted ear when compared with the contralateral control ear for 750 Hz (p < 0.01) and 1000 Hz (p < 0.05). The typical two-peak pattern of WBTA measurements was seen in both ears preoperatively but changed to a one-peak pattern in the newly implanted ear. The linear mixed model showed that not only the cochlear implantation in general but also the insertion through the round window compared to the cochleostomy leads to a decreased absorbance at 750 and 1000 Hz. Conclusions: With WBTA, we were able to detect mechanical changes of the acoustical pathway after CI surgery. The implantation of a CI led to decreased absorbance in the lower frequencies and the two-peak pattern was shifted to a one-peak pattern. The result of the linear mixed model indicates that WBTA can detect mechanical changes due to cochlear implantation not only in the middle ear but also in the inner ear.

9 February 2024 in Audiology and Neurotology
Verified
Journal Article
Open access
PDF file
CMChristian von Mitzlaff
Christian von Mitzlaff
IDIvo Dobrev
Ivo Dobrev
TFTahmine Farahmandi
Tahmine Farahmandi
FPFlurin Pfiffner
Flurin Pfiffner
CRChristof Röösli
Christof Röösli
Introduction: The mechanism of non-osseous bone conduction pathways, involving the intracranial contents (ICC) of the skull, is still not well understood. This study aimed to investigate the influence of the ICC on the skull bone wave propagation, including dependence on stimulation location and coupling. Methods: Three Thiel-embalmed whole-head cadaver specimens were studied before and after the removal of the ICC. Stimulation was via the electromagnetic actuators from commercial bone conduction hearing aids. Osseous pathways were sequentially activated by mastoid, forehead, and bone-anchored hearing aid location stimulation via a 5-Newton steel headband or percutaneously implanted screw. Non-osseous pathways were activated by stimulation on the eye and dura via a 5-Newton steel headband and a custom-made pneumatic holder, respectively. Under each test condition, the 3D motion of the superior skull bone was monitored at ∼200 points. Results: The averaged response of the skull surface showed limited differences due to the removal of the ICC. In some isolated cases, the modal pattern on the skull surface showed a trend for an upshift (∼1/2 octave) in the observed natural frequencies for drained heads. This was also consistent with an observed trend for an upshift in the transition frequency in the estimated deformation across the lateral surfaces of the temporal bones. Such changes were consistent with the expected reduction in mass and damping due to the absence of the ICC. Conclusion: Overall, the ICC affect to a limited extent the motion of the skull bone, with a limited trend for a reduction of its natural frequencies.