Background: The ablation index (AI) is reported to be useful for a durable pulmonary vein isolation (PVI). However, there have been no studies investigating the relationship between the power, contact force (CF), AI, and steam pops.Methods: Using an in vitro model, ablation energy was delivered until a steam pop occurred and the time to the steam pop and AI when the steam pop occurred were measured. The experiment was performed with a combination of various powers (20, 30, 40, and 50 W) and contact forces (CFs) (10, 30, and 50 g) 20 times for each setting. The analysis consisted of two protocols. The first protocol was a comparison between the ablation power and several parameters under the same CF (10, 30, and 50 g). The second protocol was a comparison between the CF and several parameters under the same power (20, 30, 40, and 50 W). The correlation between the lesion formation and ablation parameters was evaluated.Results: The AI value when steam pops occurred varied depending on the ablation settings. All AI median values were <500 under an ablation power of 50 W. On other hand, the median ablation time up to the steam pop was more than 46 seconds, but all median values of the AI were more than 550 under an ablation with 20 W.Conclusions: The AI cannot predict steam pops. A low power and long duration ablation could obtain a high AI value. However, high‐power ablation could not obtain a high AI value because of an early occurrence of steam pops.
Background: This study aimed to establish a systematic method for diagnosing atrioventricular nodal reentrant tachycardia (AVNRT) with a bystander concealed nodoventricular pathway (cNVP).Methods: We analyzed 13 cases of AVNRT with a bystander cNVP, 11 connected to the slow pathway (cNVP‐SP) and two to the fast pathway (cNVP‐FP), along with two cases of cNVP‐related orthodromic reciprocating tachycardia (ORT).Results: The diagnostic process was summarized in three steps. Step 1 was identification of the presence of an accessory pathway by resetting the tachycardia with delay (n = 9) and termination without atrial capture (n = 4) immediately after delivery of a His‐refractory premature ventricular contraction (PVC). Step 2 was exclusion of ORT by atrio‐His block during the tachycardia (n = 4), disappearance of the reset phenomenon after the early PVC (n = 7), or dissociation of His from the tachycardia during ventricular overdrive pacing (n = 1). Moreover, tachycardia reset/termination without the atrial capture (n = 2/2) 1 cycle after the His‐refractory PVC was specifically diagnostic. Exceptionally, the disappearance of the reset phenomenon was also observed in the two cNVP‐ORTs. Step 3 was verification of the AVN as the cNVP insertion site, evidenced by an atrial reset/block preceding the His reset/block in fast–slow AVNRT with a cNVP‐SP and slow–fast AVNRT with a cNVP‐FP or His reset preceding the atrial reset in slow–fast AVNRT with a cNVP‐SP.Conclusion: AVNRT with a bystander cNVP can be diagnosed in the three steps with few exceptions. Notably, tachycardia reset/termination without atrial capture one cycle after delivery of a His‐refractory PVC is specifically diagnostic.
1 March 2023 in Circulation: Arrhythmia and Electrophysiology