Interlaboratory comparisons of the quantized Hall resistance (QHR) are essential to verify the international coherence of primary impedance standards. Here, we report on the investigation of the stability of p-doped graphene-based QHR devices at direct and alternating currents at CMI, KRISS, and PTB. To improve the stability of the electronic transport properties of the polymer-encapsulated devices, they were shipped in an over-pressurized transport chamber. The agreement of the quantized resistance with RK/2 at direct current was on the order of 1 nΩ Ω−1 between 3.5 and 7.5 T at a temperature of 4.2 K despite changes in the carrier density during the shipping of the devices. At alternating current, the quantized resistance was realized in a double-shielded graphene Hall device. Preliminary measurements with digital impedance bridges demonstrate the good reproducibility of the quantized resistance near the frequency of 1 kHz within 0.1 μΩ Ω−1 throughout the international delivery.
6 April 2023 in Measurement Science and Technology
This paper describes the calibration of a 10 nF capacitance standard traced to the DC quantum Hall resistance primary standard using a multi-purpose digital impedance bridge. The bridge can directly compare the unlike impedances in the fully digital (FD) mode, so the capacitance standard can be calibrated by a direct comparison against the AC resistance standard. The measurement ability of the bridge in the FD mode was evaluated in various angles, and the bridge voltage dependencies were improved by selecting appropriate impedances for the high current arms and also by using an additional injection circuit. Thereby, the measurement accuracy could reach up to 10−7 in relative terms. Based on the bridge evaluation, a 10 nF capacitance standard was calibrated against a 10 kΩ resistance standard, and the measurement value agreed with a reference value in the level of 10−7.
9 May 2023 in IEEE Transactions on Instrumentation and Measurement
A digital coaxial bridge was evaluated for its frequency extension up to 100 kHz using impedance standards with known reference values. The bridge can be configured in different modes of what are known as digitally assisted (DA) or fully digital (FD) modes, and both were studied. The measurement results showed that the FD bridge cannot reach an accuracy below 100 × 10 -6 at 100 kHz, even for 1:1 comparison, unlike the DA bridge. Thus, an additional voltage injection method was attempted to achieve better accuracy for the FD bridge. As a result, 10:1 capacitance ratio measurements showed good agreement with the reference values at the level of 10 -5 for both DA and FD bridges at the highest frequency of 100 kHz.
Background: Bioelectrical impedance analysis (BIA) devices are widely used for body composition analyses, but their accuracy for body impedance measurements can vary significantly between BIA device models. Calibration of BIA device is essential to ensure the reliability of measurement data, particularly for clinical and research purposes. This study aims to evaluate the electric impedance measurement abilities of multi-frequency BIA devices using a home-built body impedance imitator (BII), designed to simulate human body impedance, and to assess the measurement uncertainty based on calibration with internationally accredited reference standards. Methods: Two different models of multi-frequency BIA device with similar specifications and measurement schemes were evaluated using the BII as a reference standard. First, the impedance values of the BII were calibrated using an LCR meter, traceable to national impedance standards. Next, the impedance measurements were performed using both models based on the standard BII over a wide frequency range (1 kHz to 1 MHz) to analyze the accuracy, linearity, contact resistance effects, etc. of the BIA devices. Results: Despite having similar specifications, two models of BIA device exhibited notable differences in their measurement performance. Model A showed better accuracy and consistency, with measurement uncertainty of approximately 1.4% at a 95% confidence level ( k = 2). In contrast, Model B showed higher deviations, particularly in trunk impedance measurements. The study found that the BIA device calibration procedure could effectively quantify the measurement uncertainty and ensure the reliability of measurement data provided by the BIA device. Conclusion: This study demonstrates that BIA devices can be calibrated using a standard procedure with a stable reference such as the BII, which traces back to internationally accredited impedance standards. The results show that the calibration significantly improves the reliability of BIA device measurements, providing a level of confidence for the measurement data.
A digitally assisted (DA) coaxial bridge, mainly consisting of a ratio transformer and digital voltage sources, was set up in a four-terminal-pair (4-TP) configuration for precision impedance ratio measurements. A 10:1 capacitance ratio was then measured in the range of 1 pF- 1 μ mainly at 1 kHz using the 4-TP DA bridge, and the results were compared to those from the existing two-terminal-pair (2-TP) and three-terminal (3-T) capacitance bridges. The deviations between the capacitance ratio measurements obtained by the different bridges lie from a few parts between 10 8 and 10 5 , depending on the measurement capabilities of the 2-TP/3-T bridges. Moreover, the measurement frequency of the 4-TP bridge was varied from 50 Hz to 20 kHz, and the frequency dependence of the 10:1 capacitance ratio was studied and compared with the reference ratios. The verification results showed that the existing 2-TP/3-T bridges can be replaced by the new 4-TP DA bridge with an extended measurement range and better uncertainty for capacitance measurements.