We report on the realization of 10 quantum Hall devices in series fabricated using epitaxial graphene on silicon carbide. Precision measurements with a resistance bridge indicate that the quantized Hall resistance across an array at a filling factor of 2 is equivalent to within the measurement uncertainty of approximately 4 10−8. A quantum Hall phase diagram for the array shows that a metrological quantization of can be achieved at the magnetic field of 6 T and a temperature of 4 K. This experiment demonstrates the possibility of timely unchangeable resistance reference in various ranges in relaxed experimental conditions.
A single electron pump was incorporated with a quantum Hall resistance and a Josephson voltage for the current evaluation in the framework of Ohm’s law. The pump current of about 60 pA level was amplified by a stable amplifier with a gain of 103 to induce a Hall voltage of about 60 mV level across a 1 MΩ Hall resistance array, which was compared with the Josephson voltage. The gain of the current amplifier was calibrated with a cryogenic current comparator bridge. For two different drive frequencies and repeated thermal cycles, the comparisons demonstrated that the pump current averaged over the first plateau was equal to ef within the combined uncertainty level of 0.3 × 10−6 (k = 1).
17 February 2019 in IEEE Transactions on Instrumentation and Measurement
A national traceability chain from dc quantum Hall resistance (QHR) to capacitance has been established at Korea Research Institute of Standards and Science using various impedance standards and coaxial bridges. An ac–dc calculable resistance standard of nominal value of 12.9 $\text{k}\Omega $ was used as a transfer standard from dc to ac at the frequency of 1.541 kHz. Two ac coaxial bridges, developed by INRiM, were employed: a ratio bridge, linking the 103- $\text{k}\Omega $ ac resistor to the 12.9- $\text{k}\Omega $ ac–dc calculable resistor calibrated in terms of the dc QHR and a quadrature bridge, calibrating the 1000 pF capacitors versus the 103- $\text{k}\Omega $ ac resistors. The ratio bridge defines the 103- $\text{k}\Omega $ resistor as a two-terminal-pair (2TP) standard and the 12.9- $\text{k}\Omega $ resistor as a four-terminal-pair standard; a cable correction was applied to the 2TP defined standard. The stability of all standards has been monitored for more than 3 years to be within a few parts in 10 6 . The capacitance value, obtained with the new traceability chain, shows an agreement (a few parts in 10 7 ) in a comparison with the reference values, traced from the National Institute for Standards and Technology (USA).
6 April 2023 in Measurement Science and Technology
This paper describes the calibration of a 10 nF capacitance standard traced to the DC quantum Hall resistance primary standard using a multi-purpose digital impedance bridge. The bridge can directly compare the unlike impedances in the fully digital (FD) mode, so the capacitance standard can be calibrated by a direct comparison against the AC resistance standard. The measurement ability of the bridge in the FD mode was evaluated in various angles, and the bridge voltage dependencies were improved by selecting appropriate impedances for the high current arms and also by using an additional injection circuit. Thereby, the measurement accuracy could reach up to 10−7 in relative terms. Based on the bridge evaluation, a 10 nF capacitance standard was calibrated against a 10 kΩ resistance standard, and the measurement value agreed with a reference value in the level of 10−7.
9 May 2023 in IEEE Transactions on Instrumentation and Measurement
A digital coaxial bridge was evaluated for its frequency extension up to 100 kHz using impedance standards with known reference values. The bridge can be configured in different modes of what are known as digitally assisted (DA) or fully digital (FD) modes, and both were studied. The measurement results showed that the FD bridge cannot reach an accuracy below 100 × 10 -6 at 100 kHz, even for 1:1 comparison, unlike the DA bridge. Thus, an additional voltage injection method was attempted to achieve better accuracy for the FD bridge. As a result, 10:1 capacitance ratio measurements showed good agreement with the reference values at the level of 10 -5 for both DA and FD bridges at the highest frequency of 100 kHz.
30 April 2021 in IEEE Transactions on Instrumentation and Measurement
A digitally assisted (DA) coaxial bridge, mainly consisting of a ratio transformer and digital voltage sources, was set up in a four-terminal-pair (4-TP) configuration for precision impedance ratio measurements. A 10:1 capacitance ratio was then measured in the range of 1 pF- 1 μ mainly at 1 kHz using the 4-TP DA bridge, and the results were compared to those from the existing two-terminal-pair (2-TP) and three-terminal (3-T) capacitance bridges. The deviations between the capacitance ratio measurements obtained by the different bridges lie from a few parts between 10 8 and 10 5 , depending on the measurement capabilities of the 2-TP/3-T bridges. Moreover, the measurement frequency of the 4-TP bridge was varied from 50 Hz to 20 kHz, and the frequency dependence of the 10:1 capacitance ratio was studied and compared with the reference ratios. The verification results showed that the existing 2-TP/3-T bridges can be replaced by the new 4-TP DA bridge with an extended measurement range and better uncertainty for capacitance measurements.