We have demonstrated a precision atom-interferometry gravimeter based on matter-wave Bragg diffraction. By decreasing the atom temperature and increasing the atom number with Raman sideband cooling, it is shown that our Bragg atom gravimeter can achieve a short-term sensitivity of ), improved by nearly a factor of 2 compared to that of a state-of-the-art atom gravimeter. The resolution of our atom gravimeter has reached after an integration time of 2000 s, comparable to a cryogenic superconducting gravimeter. Based on extremely sensitive and stable gravity measurements, a test of local Lorentz invariance in the gravity sector was performed, where the accuracy of the upper bound on the space-space component was improved by a factor of 4.
A compact laser system for a portable 87Rb atom interferometry gravimeter has been demonstrated in this work. This laser system is based on frequency doubling of a single seed laser at the wavelength of 1560 nm. The frequency of the seed laser is controlled by a digital unit with an analog feedback circuit. By using this frequency control method, the frequency of the laser system can be shifted over 1 GHz. Based on this method, the Raman frequency can be locked on the F = 3 → F′ = 4 transition of 85Rb atoms. Moreover, the Raman sideband and the repumping laser are generated by a phase modulator, and it can generate different laser frequencies to meet the requirements of a typical atom interferometer. Additional sidebands in the Raman beam produced from the phase modulator are optimized and reduced, allowing us to observe atom interference with a free evolution time of 320 ms. The control unit including the laser system has been integrated into a box with a volume of 1.5 m × 0.6 m × 0.6 m, and the weight of which is only 150 kg. Using this compact optical scheme, our atomic gravimeter has achieved a sensitivity of 53 μGal/Hz1/2 and a resolution of better than 1 μGal (1 μGal = 1 × 10−8 m/s2) in an integration time of 3000 s.
An ultra-low frequency active vibration isolator, simultaneously suppressing three-dimensional vibration noise, is demonstrated experimentally. The equivalent natural period of the isolator is 100 s and 12 s for the vertical and horizontal direction, respectively. The vibration noise in the vertical direction is about 50 times reduced during 0.2 and 2 Hz, and 5 times reduced in the other two orthogonal directions in the same frequency range. This isolator is designed for atom gravimeters, especially suitable for the gravimeter whose sensitivity is limited by vibration couplings.
As a novel instrument for measuring absolute gravity, the atom gravimeter has shown high sensitivity and stability in the past three decades. In this work, we report a car-based portable atom gravimeter whose sensitivity is at the level of (1 Gal = 0.01 m/s2) both for lab and outdoor measurements. Its resolution can reach 30 μGal within an integration time of 10 000 s when located in a quiet place. During four months of gravity observation, there is no drift for the measurement result, showing that this car-based gravimeter has a competitive advantage at long term stability. By using this car-based atom gravimeter in the open air, the field gravity measurements with sub-mGal performance have been demonstrated in the mountain gravity survey. According to the gravity measurement results from our car-based atom gravimeter, the density distribution of this mountain is inverted, which is consistent with the result from the geological exploration.
A thrust stand is developed for testing micro-Newton level thrusters on the ground. The stand is composed of a compound pendulum that is symmetrically suspended by two thin beryllium copper strips, and it is precisely calibrated by gravity. The stiffness of the stand can be adjusted in 3 orders of magnitude by a counterweight. When the stiffness is larger than 1 Nm/rad, the stand demonstrates a fast response to thrust. The measured range of the stand reaches 1000 μN, and the noise is less than 0.1 μN/ within 1 mHz–1 Hz. To calibrate the resolution of the stand, an electrostatic force is applied to the stand with an actuator. The equivalent thrust is determined to be 0.09 μN with a standard uncertainty of 0.02 μN. Using the stand, a micro-Newton colloid thruster is tested. The output of the colloid thruster changes with the applied voltage as 0.015(1) μN/V. When changing the voltage by 50 V, the change in thrust is measured to be 0.7 μN with a standard uncertainty of 0.1 μN.