We present a novel approach to fundamentally eliminate electrical feedthrough (cross-talk) signal in self-sensing piezoelectric microcantilevers, used in dynamic mode atomic force microscopy (AFM). The probe is an Si microcantilever on which a two-layer piezoelectric stack transducer is microfabricated. The top transducer in the stack functions as an actuator, while the bottom transducer is used as a deflection sensor. This arrangement leads to a collocated actuator-sensor pair that co-exist on the same surface area. We report experimental results that establish the sensor's ability to capture the full dynamic range of the cantilever with minimal feedthrough from the actuator. [2019-0143].
Active microcantilevers with on-chip sensing and actuation provide significant advantages in tapping mode Atomic Force Microscopy (AFM). Collocated transduction allows for effective manipulation of cantilever dynamics through feedback control, enabling higher scan rates. However, the adjacency of the sensing and actuation electrodes is known to result in a high level of feedthrough, leading to a low imaging resolution. Readout circuit noise further deteriorates the imaging precision. Here, we investigate the noise sources that affect AFM microcantilevers with collocated aluminum nitride (AlN) actuator–sensor pairs. We reported these cantilevers in earlier work and demonstrated that they display a very low level of feedthrough between the actuation and sensing electrodes. We present a high signal-to-noise ratio (SNR) sensing method that enables us to demonstrate high-resolution AFM on a calibration grating with nm-step silicon carbide (SiC) terraces. Measuring the Lorentzian response of the cantilever’s Brownian motion with the on-chip active sensor at resonance enables us to calibrate the dynamic stiffness at the first fundamental resonance mode, without utilizing an optical sensor.
Active Atomic Force Microscope (AFM) cantilevers with integrated actuator/sensor pairs display larger feedthrough than conventional base excited cantilevers operating in conjunction with optical displacement sensors. To control the Q factor of a cantilever, this feedthrough needs to be canceled or significantly reduced. We use an AFM microcantilever with a double-stack sensor/actuator pair to address the feedthrough cancellation problem. Our device provides two sensor signals: one used for displacement sensing and the other for velocity measurement, with the latter being used for feedback. A readout circuit is designed to sense the displacement signal and to implement the velocity feedback controller. With this method, we were able to reduce the Q factor from 390 to 30.
Thin film piezoelectric-on-silicon (TPoS) elliptical resonators are presented in this paper. The micro-resonators are composed of two rectangular resonant plates vibrating out-ofphase at the first length-extensional mode, which are coupled through arches on sides. In this way, extensional motions of rectangles are converted to rotational motions via arches which make the resonator suitable for operation under water, where all sidewalls of the resonator vibrate in parallel with liquid to maintain high liquid phase quality factors (Q). Thanks to wide electrode area available on rectangular plates, strong piezoelectric transduction is achieved leading to low liquid phase motional resistances. Different variations of the resonators were designed and fabricated to operate at different resonance frequencies. The TPoS elliptical resonators exhibit high Qs up to 7400 for operation in air while showing liquid phase Qs as high as 200. An elliptical resonator operating at 4 MHz with the lowest motional resistance of 5.6 kQ was involved in an oscillation circuit with a single-stage sustaining amplifier exhibiting frequency stability of 2 × 10 -7 (Δf/f 0 ). Such high stability makes such resonators an excellent choice for high-resolution mass or viscosity sensing at liquid phase such as biosensing applications.
31 July 2020 in Journal of Microelectromechanical Systems
This manuscript presents a novel microcantilever with an embedded piezoelectric sensor-actuator pair for dynamic atomic force microscopy (AFM). The transducer pair is constructed from a two-layered AlN stack. Stacking the piezoelectric transducers in this manner leads to a minimal feedthrough from actuation to sense electrode, granting a high dynamic range frequency response for dynamic mode AFM. The cantilever's design allows for dual mode operation at 1st and 2nd resonance modes. High resolution tapping mode imaging results are reported, while the cantilever is operated at these modes. A feedback control loop is used to modify quality factor of the 1st mode of the cantilever, using a positive position feedback (PPF) controller. A faster response time is achieved by reducing the Q-factor, enabling the cantilever to track the topography at a higher scan rate. Results of bimodal AFM imaging are reported, using amplitude changes of the 1st mode for surface topography, while material properties are encoded in phase changes of the 2nd mode.
This work presents low motional resistance piezoelectric MEMS resonators with the highest quality factor (Q) in water reported to date that can be used for direct biomarker or chemical detection in liquid media. Thin film Piezoelectric-on-substrate (TPoS) resonators operating in higher order lateral bulk acoustic modes are designed and fabricated, and their performance as mass balances in water is characterized. Q up to 312 and motional resistance as low as 662Ω were measured in water for devices with resonant frequency of 11MHz. To characterize resonator mass sensitivities, a SiO 2 passivation layer on top of one of the resonators was slowly etched in several consecutive steps and frequency shifts were recorded showing of 33.4 cm 2 /ng. Higher order mode operation and relatively large size of the devices (high volume/surface ratio) along with wide gaps between the resonating body and the substrate are considered to be the main factors minimizing viscous damping in liquid.