Non-contact tonometers based on the method using air puff and Scheimpflug’s fast camera are one of the latest devices allowing the measurement of intraocular pressure and additional biomechanical parameters of the cornea. Biomechanical features significantly affect changes in intraocular pressure values, as well as their changes, may indicate the possibility of corneal ectasia. This work presents the latest and already known biomechanical parameters available in the new offered software. The authors focused on their practical application and the diagnostic credibility indicated in the literature. An overview of available literature indicates the importance of new dynamic corneal parameters. The latest parameters developed on the basis of biomechanics analysis of corneal deformation process, available in non-contact tonometers using Scheimpflug’s fast camera, are used in the evaluation of laser refractive surgery procedures, e.g. LASIK procedure. In addition, the assessment of changes in biomechanically corrected intraocular pressure confirms its independence from changes in the corneal biomechanics which may allow an intraocular pressure real assessment. The newly developed Corvis Biomechanical Index combined with the corneal tomography and topography assessment is an important aid in the classification of patients with keratoconus. New parameters characterising corneal deformation, including Corvis Biomechanical Index and biomechanical compensated intraocular pressure, significantly extend the diagnostic capabilities of this device and may be helpful in assessing corneal diseases of the eye. Nevertheless, further research is needed to confirm their diagnostic pertinence.
Background: The Corvis® ST tonometer is an innovative device which, by combining a classic non-contact tonometer with an ultra-fast Scheimpflug camera, provides a number of parameters allowing for the assessment of corneal biomechanics. The acquired biomechanical parameters improve medical diagnosis of selected eye diseases. One of the key elements in biomechanical measurements is the correct corneal contour detection, which is the basis for further calculations. The presented study deals with the problem of outer corneal edge detection based on a series of images from the afore-mentioned device. Corneal contour detection is the first and extremely important stage in the acquisition and analysis of corneal dynamic parameters. Result: A total of 15,400 images from the Corvis® ST tonometer acquired from 110 patients undergoing routine ophthalmologic examinations were analysed. A method of outer corneal edge detection on the basis of a series of images from the Corvis® ST was proposed. The method was compared with known and commonly used edge detectors: Sobel, Roberts, and Canny operators, as well as others, known from the literature. The analysis was carried out in MATLAB® version 9.0.0.341360 (R2016a) with the Image Processing Toolbox (version 9.4) and the Neural Network Toolbox (version 9.0). The method presented in this paper provided the smallest values of the mean error (0.16%), stability (standard deviation 0.19%) and resistance to noise, characteristic for Corvis® ST tonometry tests, compared to the methods known from the literature. The errors were 5.78 ± 9.19%, 3.43 ± 6.21%, and 1.26 ± 3.11% for the Roberts, Sobel, and Canny methods, respectively. Conclusions: The proposed new method for detecting the outer corneal contour increases the accuracy of intraocular pressure measurements. It can be used to analyse dynamic parameters of the cornea.