Purpose: To validate a previously proposed algorithm that modifies a mammogram to appear as if it was acquired with different technique factors using realistic phantom-based mammograms. Approach: Two digital mammography systems (an indirect- and a direct-detector-based system) were used to acquire realistic mammographic images of five 3D-printed breast phantoms with the technique factors selected by the automatic exposure control and at various other conditions (denoted by the original images). Additional images under other simulated conditions were also acquired: higher or lower tube voltages, different anode/filter combinations, or lower tube current–time products (target images). The signal and noise in the original images were modified to simulate the target images (simulated images). The accuracy of the image modification algorithm was validated by comparing the target and simulated images using the local mean, local standard deviation (SD), local variance, and power spectra (PS) of the image signals. The absolute relative percent error between the target and simulated images for each parameter was calculated at each sub-region of interest (local parameters) and frequency (PS), and then averaged. Results: The local mean signal, local SD, local variance, and PS of the target and simulated images were very similar, with a relative percent error of 5.5%, 3.8%, 7.8%, and 4.4% (indirect system), respectively, and of 3.7%, 3.8%, 7.7%, and 7.5% (direct system), respectively. Conclusions: The algorithm is appropriate for simulating different technique factors. Therefore, it can be used in various studies, for instance to evaluate the impact of technique factors in cancer detection using clinical images.
To validate a candidate instrument, to be used by different professionals to assess image quality in digital mammography (DM), against detection performance results. A receiver operating characteristics (ROC) study was conducted to assess the detection performance in DM images with four different image quality levels due to different quality issues. Fourteen expert breast radiologists from five countries assessed a set of 80 DM cases, containing 60 lesions (40 cancers, 20 benign findings) and 20 normal cases. A visual grading analysis (VGA) study using a previously-described candidate instrument was conducted to evaluate a subset of 25 of the images used in the ROC study. Eight radiologists that had participated in the ROC study, and seven expert breast-imaging physicists, evaluated this subset. The VGA score (VGAS) and the ROC and visual grading characteristics (VGC) areas under the curve (AUCROC and AUCVGC) were compared. No large differences in image quality among the four levels were detected by either ROC or VGA studies. However, the ranking of the four levels was consistent: level 1 (partial AUCROC: 0.070, VGAS: 6.77) performed better than levels 2 (0.066, 6.15), 3 (0.061, 5.82), and 4 (0.062, 5.37). Similarity between radiologists' and physicists' assessments was found (average VGAS difference of 10 %). The results from the candidate instrument were found to correlate with those from ROC analysis, when used by either observer group. Therefore, it may be used by different professionals, such as radiologists, radiographers, and physicists, to assess clinically-relevant image quality variations in DM.
The aim of this work is the generation of realistic synthetic mammograms, using as an input of the imaging acquisition simulation process digital anthropomorphic phantoms, reconstructed from sets of dedicated breast computed tomography (BCT) images from different patients. The voxel-based structure and the segmentation into fibroglandular, adipose and skin tissues are performed through trivariate tensor-product B-spline approximation and morphological operations. The obtained phantoms can be modified by means of geometrical transformations that replicate typical breast shape deformities, and by locally introducing virtual masses and calcifications. After simulating biomechanical compression of the 3D breast phantoms, we generate the mammograms in both craniocaudal (CC) and mediolateral oblique (MLO) views, modelling the x-ray interaction with breast tissues with a Monte Carlo approach implemented in the in silico breast imaging pipeline VICTRE. The methodology proposed here can contribute to the creation of synthetic mammogram databases, to be used for in silico testing of diagnostic and therapeutic techniques, as well as for the validation of artificial intelligence (AI) systems in diagnostic imaging and cancer screening. The great advantage is that, from a single BCT scan, it is possible to generate multiple realistic mammograms, with different anatomical features, in terms of breast shape and size, and type and location of lesions.