A white light non-scanning correlation interrogation system was proposed and built to interrogate absolute length of the air cavity of fiber-optic compound Fabry–Perot pressure sensors for the extraction of pressure value. By carefully choosing thickness range and tilt angle of the optical wedge used for cavity length matching, correlation interferometric signal of the basal cavity can be naturally filtered out. Based on peak positioning by Fourier transform, bandpass filtering in frequency domain, inverse Fourier transform back to time domain, envelope fitting and zero fringe finding through a gravity center method, cavity length can be determined with an accuracy of 0.04%. The system was used for the interrogation of a fiber-optic compound Fabry–Perot pressure sensor under different pressures. For a pressure range of 0.1~2.9 Mpa, the linear relationship between the air cavity length and the gas pressure imposed was successfully extracted.
This study presents an extrinsic Fabry–Perot interferometric (EFPI) fiber-optic strain sensor with a very short cavity. The sensor consists of two vertically cut standard single-mode fibers (SMFs) and a glass capillary with a length of several centimeters. The two SMFs penetrate into the glass capillary and are fixed at its two ends with the use of ultraviolet (UV) curable adhesives. Based on the use of the lengthy glass capillary sensitive element, the strain sensitivity can be greatly enhanced. Experiments showed that the microcavity EPFI strain sensor with initial cavity lengths of 20 μm, 30 μm, and 40 μm, and a capillary length of 40 mm, can yield respective cavity length–strain sensitivities of 15.928 nm/με, 25.281 nm/με, and 40.178 nm/με, while its linearity was very close to unity for strain measurements spanning a range in excess of 3500 με. Furthermore, the strain–temperature cross-sensitivity was extremely low.
To solve the cavity interrogation problem of short cavity fiber Fabry–Perot sensors in white light spectral interrogation with amplified spontaneous emissions (ASEs) as the white light sources, a data processing method, using an improved elliptical fitting equation with only two undetermined coefficients, is proposed. Based on the method, the cavity length of a fiber Fabry–Perot sensor without a complete reflection spectrum period in the frequency domain can be interrogated with relatively high resolution. Extrinsic fiber Fabry–Perot air-gap sensors with cavity lengths less than 30 μm are used to experimentally verify the method, and are successfully interrogated with an accuracy better than 0.55%.
A fiber-optic temperature sensor based on fiber tip polystyrene microsphere is proposed. The sensor structure can be formed simply by placing and fixing a polystyrene microsphere on the center of an optical fiber tip. Since polystyrene has a much larger thermal expansivity, the structure can be used for high-sensitive temperature measurement. By the illuminating of the sensor with a broadband light source and through the optical Fabry-Perot interference between the front and back surfaces of the polystyrene microsphere, the optical phase difference (OPD) or wavelength shift can be used for the extraction of temperature. Temperature measurement experiment shows that, using a fiber probe polystyrene microsphere temperature sensor with a spherical diameter of about 91.7 µm, a high OPD-temperature sensitivity of about −0.617 96 nm/°C and a good linearity of 0.991 6 were achieved in a temperature range of 20°C–70°C.
9 October 2024 in IEEE Transactions on Instrumentation and Measurement
Recent developments of various types of fiber-optic vector magnetic field sensors based on magnetic fluid are reviewed. From the perspective of sensing principles, these senors are classified into Mach-Zehnder interferometer sensor, surface plasmon resonance sensor, Fabry-Perot interferometer sensor, fiber grating sensor, Sagnac interferometer sensor and Michaelson interferometer sensor. Working mechanisms, fabrication methods, and performances of these sensors are discussed in details respectively. Finally, these sensors are compared and analyzed from measurement range, sensitivity, mechanical strength, etc. Some current common problems and prospects for further development are pointed out. Furthermore, this work gives some suggestions to improve the sensitivity and mechanical strength, which can also be applied in fields such as biomedical testing and undersea environmental monitoring where measurement accuracy and mechanical strength are both required.