Pelvic implants require complex geometrical shapes to reconstruct unusual areas of bone defects, as well as a high mechanical strength in order to endure high compressive loads. The electron beam melting (EBM) method is capable of directly fabricating complex metallic structures and shapes based on digital models. Fixation design is important during the 3D printing of pelvic implants, given that the fixation secures the pelvic implants to the remaining bones, while also bearing large amounts of the loads placed on the bone. In this study, a horseshoe-shaped plate fixation with a bridge component between two straight plates is designed to enhance the mechanical stability of pelvic implants. The aim of this study is to investigate the biomechanics of the horseshoe-shaped plate fixation in a 3D-printed pelvic implant using a finite element (FE) simulation. First, computed tomography (CT) scans were acquired from a patient with periacetabular bone tumors. Second, 3D FE implant models were created using the patient’s Digital Imaging and Communications in Medicine (DICOM) data. Third, a FE simulation was conducted and the stress distribution between a conventional straight-type plate model, and the horseshoe-shaped plate model was compared. In both of the models, high-stress regions were observed at the iliac fixation area. In contrast, minimal stress regions were located at the pubic ramus and ischium fixation area. The key finding of this study was that the maximal stress of the horseshoe-shaped plate model (38.6 MPa) was 21% lower than that of the straight-type plate model (48.9 MPa) in the iliac fixation area. The clinical potential for the application of the horseshoe-shaped plate fixation model to the pelvic implant has been demonstrated, although this is a pilot study.
Purpose: Autogenous arteriovenous fistulas are the preferred access for hemodialysis. Yet when created, fistulas often fail to mature, requiring surgical or radiologic interventions before their use. This pilot study measures the vascular wall elasticity and flow gradient using an open-source ultrasound software program designed to aid in assessing fistula maturation. Methods: A total of seven end-stage renal disease patients were enrolled for our study after providing informed consent. Ultrasound scanning was performed for the inflow artery, post-arterial anastomotic segment, and outflow vein at initial and follow-up evaluation. Conventional digital imaging and communications in medicine data were collected from the ultrasound machine. The vessel diameter and distensibility of artery, post-arterial anastomotic segment, and vein were computed from the digital imaging and communications in medicine data using an open-source ultrasound software program. Results: The vessel diameter of artery and vein increased from 4.6 ± 1.1 mm to 6.0 ± 1.1 mm and from 5.8 ± 0.7 mm to 7.5 ± 0.9 mm from 1 to 6 weeks post-operation, respectively. Conversely, the vessel diameter of post-arterial anastomotic segment decreased from 4.2 ± 1.0 mm to 3.5 ± 0.9 mm from 1 to 6 weeks post-operation. The distensibility of artery and post-arterial anastomotic segment increased from 3.4% ± 0% to 5.9% ± 1.1% and 3.7% ± 1.2% to 4.9% ± 1.4%, respectively, while the distensibility of vein decreased from 5.0% ± 1.3% to 2.6% ± 0.4% from 1 to 6 weeks post-operation. Conclusion: This study demonstrates that the change in vessel diameter and distensibility related to the healthy remodeling as the vein dilates during maturation.
Vascular access is essential for hemodialysis patients. The mature native arteriovenous fistula has been the preferred vascular access for hemodialysis, because it has greater longevity than synthetic grafts. However, once surgically created, fistulas often fail to develop (mature) into viable points of vascular access, requiring surgical or radiologic interventions before their use. Because maturation depends on vascular mechanics (e.g., distensibility and wall shear), we developed open-source ultrasound software to investigate these metrics clinically. We demonstrated in a single patient the ability of the software for consistent measurements from various locations within a cardiac cycle and between different cardiac cycles. We further assessed the ability of the software to identify changes in distensibility of a patient’s fistula from 1 to 6 weeks postoperation. The routine frame rates of clinical machines demonstrated high fidelity tracking within cardiac cycles (coefficient of variation [CV] = 2.4% ± 0.011) and between cardiac cycles (CV = 2.4% ± 0.004). The distensibility of the patient’s fistula from 1 to 6 weeks postoperation increased from 4% to 7% in the arterial inflow and from 3% to 4% in the postarterial anastomotic segment (PAAS). In contrast, the distensibility of the outflow vein decreased from 4% to 2%. These results corroborate that in addition to diameter changes, the mechanical properties of the vascular segments changed during fistula maturation. This demonstrates that our software-based approach may allow ultrasound-based mechanical measurements to become more accessible for wider clinical research.