Nils Weinrich

Dr. Nils Weinrich


Research Focus & Metrics

Academic Career Length

8Years (2019-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
CAChristian Apelmann
Christian Apelmann
BKBirgitt Kowald
Birgitt Kowald
NWNils Weinrich
Nils Weinrich
JDJens Dischinger
Jens Dischinger
Albert NienhausAlbert Nienhaus
Albert Nienhaus
CJChristian Jürgens
Christian Jürgens
Background: Due to the drastic reduction of the eye lens dose limit from 150 mSv per year to 20 mSv per year since 2018, the prospective investigation of the estimated dose of the eye lens by radiological imaging procedures at the surgical site during trauma surgery in the daily work process was carried out. This was also necessary because, as experience shows, with changes in surgical techniques, there are also changes in the use of radiological procedures, and thus an up-to-date inventory can provide valuable information for the assessment of occupationally induced radiation exposure of surgical personnel under the current conditions. Methods: The eye lens radiation exposure was measured over three months for five trauma surgeons, four hand surgeons and four surgical assistants with personalized LPS-TLD-TD 07 partial body dosimeters Hp (0.07). A reference dosimeter was deposited at the surgery changing room. The dosimeters were sent to the LPS (Landesanstalt für Personendosimetrie und Strahlenschutzausbildung) measuring institute (National Institute for Personal Dosimetry and Radiation Protection Training, Berlin) for evaluation after 3 months. The duration of the operation, occupation (assistant, surgeon, etc.), type of surgery (procedure, diagnosis), designation of the X-ray unit, total duration of radiation exposure per operation and dose area product per operation were recorded. Results: Both the evaluation of the dosimeters by the trauma surgeons and the evaluation of the dosimeters by the hand surgeons and the surgical assistants revealed no significant radiation exposure of the eye lens in comparison to the respective measured reference dosimeters. Conclusions: Despite the drastic reduction of the eye lens dose limit from 150 mSv per year to 20 mSv per year, the limit for orthopedic, trauma and hand surgery operations is well below the limit in this setting.

16 September 2025 in Sensors
Verified
Journal Article
Open access
PDF file
ASArndt Peter Schulz
Arndt Peter Schulz
BKBirgitt Kowald
Birgitt Kowald
MMMuench Muench
Muench Muench
KSKlaus Seide
Klaus Seide
NWNils Weinrich
Nils Weinrich
BKBenjamin Kienast
Benjamin Kienast
A total of 66 smart implants were included. As a measure of bony stability, the relative elastic compliance of the osteosynthesis was determined from the gradient between the applied external load and the measured implant load over the entire healing process. The healing process of non-unions of the femur with a smart implant was tracked by telemetric measurements over a timespan of up to 10 years. The measurements of the longest healing process show a very slow but constant decrease in force transmission over the implant, radiological findings over 10 years show corresponding consolidation until bony healing. The use of a telemetrically instrumented bone plate, a so-called smart implant, to monitor the healing process is a successful procedure to support the clinician in his decision to take further surgical measures or to wait until healing occurs.