This paper describes investigation of the influence of the protective layer thickness on the calibration sensitivity of PVDF films sensors for the cavitation impacts measurements. The PVDF film sensor is casted into an aluminium block. The drop ball method is used for the measurement of the relation between impact force and the voltage detected on the PVDF film sensor. The calibration constants are measured for three different protective layers thicknesses. Five different ball weights for 400 mm drop height are used to reach the required impact force range. The ball positions for the evaluation of the impact force are measured with a high speed camera. The voltage signal detected on the PVDF film clamps was measured with a high speed digitizer. The measured signals are analysed in LabVIEW Signal Express.
A PVDF pressure sensor was used to measure the pressure peaks due to the collapse of cavitation bubbles in the high-speed tunnel of the LEGI laboratory. It was flush mounted on a stainless steel disk in the most erosive area of the test section. The recorded data were post- processed in order to get the impact load spectra for different velocities at constant cavitation number. The results are presented as cumulative histograms of peak rate and maximal impact load for different flow velocities of the high-speed tunnel.