In the present work, an attempt has been made to identify an organic compound that can perform as a secondary calibrator of a Coupled-Optical Fiber Raman Spectrograph (COFRS) at 785 nm in the range of 100–3300 cm −1 . Considering not only the chemical stability and purity but also the uneven spectral distribution of Raman peaks of some organic molecules such as Benzoic Acid, Acetone, Benzene, Methylene Chloride, and Dimethyl Sulfoxide, a wavenumber calibration analysis was developed. Accordingly, the wavenumber calibration behavior was monitored by obtaining Raman spectra of diverse standards with different functional groups, such as Ethanol, Methanol, 2-Chlorophenol, 4-Chlorophenol, Chloroform, 1,2,4-Triazole, Cyclohexanone, n-Hexane, Toluene, Sodium Dodecyl Sulfate, and Ethyl Acetate. Since acetonitrile is used as a COFRS calibrator with several tall peaks throughout the spectral range—a preferable feature for obtaining an acceptable calibration—it has been found that there is a low dispersion of the central wavenumber maxima in the Raman signals for Dimethyl Sulfoxide as well as Benzene, Acetone, Benzoic Acid, and Methylene Chloride during the wavenumber calibration in both the entire and fingerprint spectral ranges of COFRS, respectively. This result demonstrates that COFRS can be calibrated with Dimethyl Sulfoxide, despite its uneven spectral distribution of Raman peaks along the analyzed spectral range.