1 October 2019 in IEEE Transactions on Biomedical Circuits and Systems
An electrical impedance tomography (EIT) system based on frequency division multiplexing (FDM) is proposed for real-time lung physiological imaging. The FDM technique allows the integration of 13 dedicated voltage sensing channels by combining data on-chip and sharing of ADC to alleviate area penalty caused by multi-channel. The EIT SoC is of the following features: 1) early I/Q demodulation to relax the bandwidth requirement of analog front end and minimize the impact of motion artifacts and DC electrode offset; 2) eliminates the need of adaptive gain control (AGC) with constant inverted “V-shape” gain configuration to compensate amplitude variations across all channels; 3) FDM to combine 13 pairs of I/Q signals into two data streams for quantization using only two Modulators (DSMs); 4) batch data recovery by Blackman window corrected FFT without any digital filtering involved; and 5) lowest power consumption and smallest area occupation per channel reported to date. The EIT SoC occupies an area of 11.28-mm2 in 130nm CMOS technology with a total power consumption of 1.53-mW under 1-V power supply. As a result, it generates lung EIT images at up to 5 frames per second (fps).
26 March 2019 in IEEE Journal of Solid-State Circuits
A dual-loop eight-channel electrocardiogram (ECG) recording system with the fast settling mode is proposed for wireless real-time 12-lead ECG applications. It employs the frequency division multiplexing (FDM) technique to achieve multi-channel IA (MCIA). The FDM technique allows using smaller input capacitor and sharing of programmable gain amplifier (PGA), buffer, and ADC to achieve smaller area. To suppress the dc offset of bio-signal with faster settling, dual dc servo loops are proposed. The main loop employs the least-mean-square (LMS) algorithm in the digital domain to extract the dc offsets in different frequency channels. An auxiliary fast settling loop (FSL) in the analog domain can be enabled to shorten the settling time from 14 s down to only 0.6 s. A 16-bit Σ Δ ADC, including the decimation filter, is adopted to achieve high resolution and avoid using high Q anti-aliasing filter in the analog domain. The MCIA achieves 112.5-dB common mode rejection ratio (CMRR), 113.0-dB power supply rejection ratio (PSRR), and 1.44-μ Vrms noise performance. The MCIA occupies only 0.3 mm² and consumes 60 μW for eight channels. The whole ECG recording system was implemented in 130-nm CMOS technology.
15 December 2022 in IEEE Transactions on Biomedical Circuits and Systems
This paper proposes and experimentally validates a novel concentric circle (CC) model for indoor WiFi sensing. By setting the transmitter and receiver together, the perception model becomes concentric circles with equal spacing, eliminating the blind zone and unequal radial sensitivity problems of the Fresnel zone (FZ) model. Then a human respiratory monitoring system is developed based on this model, which executes the following steps: (1) Principal component analysis (PCA) is applied to the channel state information ratio (CSIR) as a preprocessing to extract the components related to human activities. (2) Human presence and respiratory signal detection are adopted to improve monitoring accuracy. (3) The Doppler respiratory frequency is extracted to calculate the respiratory rate. Experimental results show that the CC model achieves high accuracy in velocity measurement with an error of less than 0.4 cm/s. The respiration monitoring system can accurately monitor human respiration with an error of less than 0.7 bpm within 6 m.