17 July 2019 in IEEE Electron Device Letters
We present a novel silicon photomultiplier based on a shallow junction obtained within a standard 0.35 ${\mu }\text{m}$ complementary metal oxide semiconductor process. The sensitive structure is equipped with a dedicated optical window as well as an optimized antireflective coating. The device exhibits performances comparable to custom-based commercial technologies: an unprecedented photon detection efficiency of 43% at 420 nm, a breakdown voltage temperature dependence of (27.9 ± 0.9) mV/K, a dark count rate of 480 kHz/mm 2 , a gain of ${3\times 10^{6}}$ , and a single photon time resolution of (78 ± 2) ps (FWHM). The process proposed in this letter can be applied to new generation single photon sensors with on-chip integrated intelligent electronics.
23 January 2024 in IEEE Transactions on Radiation and Plasma Medical Sciences
Novel design solutions for dedicated portable brain positron emission tomography systems with improved performance facilitate emergency and interventional image-guided surgery as well as advanced diagnostics of clinically impaired patients with neurodegenerative diseases. We report a novel portable MVT-based All-Digital helmet PET system with a hemispherical detector arrangement based on the Multi Voltage Threshold technology. It has a transverse and axial field of view (FOV) of 200 mm and 124 mm, respectively. It allows to scan subjects in a standing, sitting, and lying position. We evaluated the performance of the system according to NEMA standards. The scanner exhibits a noise equivalent count rate peak of (151±2) kcps at the activity of 40.65 kBq/mL, a sensitivity of (55.24±0.05) cps/kBq, and a spatial resolution at the center of the FOV of approximately 3.3 mm (FWHM), when using the filtered back projection algorithm. For a mini Derenzo phantom, rods of 2.0 mm diameter can be clearly separated. Time-dynamic 18F-Fluorodeoxyglucose human brain imaging was performed, showing the distinctive traits of cortex and thalamus uptake, as well as of the arterial and venous flow with 30 s time frames. We finally verified the usability of the device in the diagnostics of Alzheimer’s Disease by imaging human subjects with 18F-Florbetapir.
16 January 2025 in IEEE Transactions on Electron Devices
Silicon photomultipliers (SiPMs) fabricated using standard CMOS processes offer the advantages of compatibility with on-chip digital circuits, making them promising candidates among low-photon flux sensors. However, the high noise and low-photon detection efficiency (PDE) of CMOS SiPMs have limited their widespread adoption. Here, we present a novel SiPM obtained within a standard 0.11- $\mu $ m CMOS process. We report how guard ring (GR) width and distance from shallow trench isolation (STI) may affect the noise performance, and we optimize for them. The final device demonstrates performance that is competitive with both custom-based commercial technologies and other CMOS-based SiPMs, featuring an unprecedented peak PDE of 62% at 420 nm, a dark count rate (DCR) of $\mathbf {\text {(}{174}\pm {7}\text {)}}$ kcps/mm $\mathbf {^{{2}}}$ , a gain of $\mathbf {\text {(}{2}.{0}\pm {0}.{2}\text {)}\times {10}^{{6}}}$ and a single-photon time resolution (SPTR) of $\mathbf {\text {(}{74}\pm {3} \text {)}}$ ps [full width at half maximum (FWHM)] at an excess bias voltage of 3 V.
28 November 2023 in IEEE Transactions on Nuclear Science
Positron Emission Tomography (PET) is increasingly employed for precision monitoring in plant science, as it is a non-invasive technique for the quantitative analysis of the functional mechanisms of plant metabolism. When exported to digital agronomy in combination with qualitative detection, PET has the potential to drive the sustainable management of chemical fertilizers. Such a translational approach requires a transition from controlled in-lab to natural in-field PET imaging. We report a novel portable and shape-adaptable digital plant PET system based on the Multi Voltage Threshold digital readout, with an extendable transverse Field of View (FOV) ranging between 80 mm and 120 mm. The system reaches a peak sensitivity up to (16.94 ± 0.01)% and (9.12 ± 0.01)% in closed and maximally opened FOV configuration, respectively. Image quality has been tested with a dedicated plant phantom imaging. We verified the ability of the prototype to perform plant PET imaging: in-lab with Zea mays L . sprouts under heat stress, and in-field with Triticum aestivum sprouts under drougt stress.
7 July 2025 in Physics in Medicine & Biology
Objective. Plant Positron Emission Tomography (PET) is a new and efficient imaging technique which aims at providing a quantitative analysis of plant stress, enabling personalized crop management and maximizing productivity. However, a highly performant classification system for noisy dynamic plant PET images faces the challenge of retrieving noise-free datasets and encoding both spatial and temporal representations within a unified model. Approach. To overcome these limitations, we introduce an innovative hybrid model that combines denoising and classification for dynamic plant PET images. Initially, we compute a precise solution for the denoising problem of noisy dynamic plant PET images using a modified optimization method coupled with deep convolutional neural networks. Subsequently, this solution is unfolded into a deep network known as the predictor–corrector network (PCNet). To optimize the PCNet without requiring a noise-free dynamic training set, we propose a novel unsupervised learning method. Finally, the sequence of noise-reduced dynamic plant PET images is further fed into a unique classification system, encoding spatial representations of images and temporal representations of multivariate time series into a unified spatiotemporal representation and generating a prediction. Main results. The experimental results underscore the necessity of the denoising procedure and highlight the superiority of the proposed PCNet over existing competing denoising methods, demonstrating the effectiveness of the proposed classification system. Notably, the classification performance between the two classes achieves an averaged accuracy of 0.852, an averaged precision of 0.838, an averaged recall of 0.959, and an averaged F1-score of 0.880. Significance. The ability of the proposed method to reduce noise intensity and effectively encode spatiotemporal representations overcomes the limitations of existing methods. This advancement may have substantial implications for other noisy dynamic image classification.
11 December 2024 in IEEE Transactions on Radiation and Plasma Medical Sciences
Risk stratification algorithms for gastrointestinal stromal tumors (GISTs) are mainly based on Computed Tomography (CT) data. Though 18F-Fluorodeoxyglucose Positron Emission Tomography (18F-FDG PET) imaging may improve their performance, challenges in image interpretation in the gastrointestinal tract still limit the widespread integration of PET into routine clinical protocols, causing a poor availability of PET data to develop and train stratification models. To solve this issue, we propose to enrich existing 18F-FDG PET GIST datasets with pseudo-images generated with a novel Conditional PET Generative Adversarial Network (CPGAN), which employs a weighted fusion of CT images and tumor masks, embedding also clinical data. As for GIST assessment, we propose the Transformer-Based Multi-Modal Network for GIST Risk Stratification (TMGRS), which is trained on the enriched dataset and exploits the properties of transformers to process simultaneously PET and CT images. The training and validation of the models were conducted on a multi-center dataset comprising 208 patients. In comparison with the existing stratification methods, CPGAN-synthesized PET images show a Peak Signal-to-Noise Ratio increased on average by 18% and improve risk stratification, which achieves a remarkable accuracy of 0.937 when TMGRS network is used. Results underscore the potential of CPGAN network in providing more reliable GIST predictions.