This study aimed to present and validate the Autonomous and Intelligent Packaging for the Cold Chain in Health Systems (EMAIS‐SR) for organ transport through an experimental model using porcine organs. To do so, Landrace porcine hearts, livers, pancreases and kidneys were transported using either EMAIS‐SR or traditional ice‐filled coolers. Following transport, organs were assessed after 2 h through collected data that included temperature monitoring, macroscopic evaluation, histological analysis and microbiological testing. Prototypes at different technology readiness levels (TRL6 and TRL7) were evaluated. Regarding the findings, EMAIS‐SR maintained organ temperatures predominantly within the recommended 2°C–8°C range. In the adjusted analysis, the TRL7 prototype achieved significantly lower post‐transport temperatures than TRL6 (adjusted difference: 2.91°C; 95% CI, 1.32–4.50; p = 0.0006), indicating improved thermal performance after system refinements. Additionally, minor macroscopic changes were observed in some samples, but without histological correlation. Histological analysis showed better tissue preservation in EMAIS‐SR samples, particularly in nuclei morphology and cytoplasmic structure. Furthermore, microbiological results indicated equivalent or lower contamination in EMAIS‐SR compared to the control, especially in later tests. Overall, EMAIS‐SR demonstrated effective thermal control, preservation of tissue integrity and biological safety in a structured preclinical model using porcine organs. Consequently, these findings validate EMAIS‐SR as a robust and clinically promising alternative to conventional organ transport systems.
BACKGROUND: Intermediate temperatures (10°C) better preserve mitochondrial function and extend ischemic time, which are fundamental for lung preservation during transplantation. The EMAIS-SR device was developed as a reusable, cost-effective alternative to maintain the 10°C set-point, thereby overcoming logistical and economic barriers in organ transport. OBJECTIVE: This study aimed to validate the thermal stability of the EMAIS-SR system in a swine model by evaluating the organ-cooling curve and the time spent within the target temperature range. DESIGN AND SETTING: Validation study using experimental surgery on porcine organs. METHODS: Lung blocks were harvested from two Landrace swine in independent experiments (P1, P2). The organs were placed in triple sterile packaging. Thermal stability was monitored for 48 h by using four negative temperature coefficient (NTC) thermistors (three intrapulmonary and one extrapulmonary) with data acquisition at a 0.1-s interval. Another datalogger (Hantek ® HTM208B) was used for control and validation. Thermal distribution was further characterized by infrared thermography (FLIR C3-X). RESULTS: In Experiment 1 (P1), intrapulmonary temperatures reached the predefined target range (≤ 12°C) approximately 1.4 h after lung extraction, and the mean temperature recorded in the container was 8.01 °C (standard deviation, 2.88°C). In Experiment 2 (P2), a mean temperature reduction of 6.6°C was observed, bringing the intrapulmonary temperature to the target range (≤ 12°C) within approximately 2 h post-extraction. CONCLUSION: The EMAIS-SR device effectively maintained lung-preservation temperatures within the target 8°C–12°C range over 48 h. These results demonstrate the system’s technical feasibility and its capacity to provide consistent thermal stability across intrapulmonary compartments.