Efficient and precise measurements of optics with parallel surfaces are crucial for ensuring component performance. However, current interferometry faces challenges from uncertain number of interference sources and adverse effects on parameter demodulation due to sampling errors and noise. To address these issues, we first use the MDL criterion on a Hankel matrix of intensity signals to estimate the number of effective surfaces by minimizing an information-theoretic cost function. Then, VMD iteratively extracts the superimposed signal into IMCs with distinct frequencies. Frequency estimation employs a zero-padded Fourier transform with a Blackman window, followed by phase and surface reconstruction. In error analysis, under SNR = 30 dB~50 dB and up to 20% phase-shifting errors, frequency solution errors remain below 0.7%. The Monte Carlo trials show low mean and standard deviation in solution errors. Repeated measurements on a transparent plate yield stable profiles with minimal PV, RMS, and PVQ errors.