Key Laboratory of Ultra-Weak Magnetic Field Measurement Technology, Ministry of Education, School of Instrumentation Science and Optoelectronics Engineering, Beihang University, Beijing 100191, China
Auto-generated profile, based on publicly available information
Teng Li
Key Laboratory of Ultra-Weak Magnetic Field Measurement Technology, Ministry of Education, School of Instrumentation Science and Optoelectronics Engineering, Beihang University, Beijing 100191, China
This paper presents a design method of biplanar coils to realize precise compensation for residual magnetic fields inside magnetic shielding rooms (MSR) based on an improved particle swarm optimization algorithm with the moore-penrose pseudoinverse. The mirror image method (IM) is introduced into the traditional target field method (TFM) to weaken the coupling effect between the magnetic shielding material and the magnetic field generated by the coil. An improved particle swarm optimization (IPSO) algorithm with the moore-penrose pseudoinverse is used to obtain the coils’ parameters, thus determining the shapes of the coils. The simulation results show that the maximum deviations of Bx and Bz coil designed by this method are improved by 74.7% and 59.5%, respectively.
Direct current (DC) bias induced by the DC transmission and geomagnetically induced current is a critical factor in the abnormal operation of electrical equipment and is widely used in the field of power transmission and distribution system state evaluation. As the main affected component, the vector magnetization state of a transformer core under DC bias has rarely been studied, resulting in inaccurate transformer operation state estimations. In this paper, a dynamic vector hysteresis model that considers the impact of rotating and DC-biased fields is introduced into the numerical analysis to simulate the distribution of magnetic properties, iron loss and temperature of the transformer core model and a physical 110 kV single-phase autotransformer core. The maximum values of B, H and iron loss exist at the corners and T-joint of the core under rotating and DC-biased fields. The corresponding maximum value of the temperature increase is found in the main core limb area. The temperature rise of the 110 kV transformer core under various DC-biased conditions is measured and compared with the FEM (Finite Element Method) results of the proposed model and the model solely based on the magnetization curve B||H. The calculation error of the temperature rise obtained by the improved model is approximately 3.76–15.73% and is much less than the model solely based on magnetization curve B||H (approximately 50.71–66.92%).
To achieve a near-zero magnetic field environment, the use of permalloy sheets with high-performance magnetic properties is essential. However, mainstream welding processes for magnetically shielded rooms (MSRs), such as argon arc welding and laser welding, can degrade the magnetic properties of the material. Additionally, neglecting the anisotropy of permalloy sheets can introduce unpredictable errors in the evaluation of MSR performance. To address this issue, this paper proposes a modified model for calculating the shielding factor (SF) of MSRs that incorporates the anisotropic magnetic characteristics of permalloy sheets. These characteristics were measured using a two-dimensional single sheet tester (2D-SST). A high-precision measurement system was developed, comprising a 2D-SST (to generate two-dimensional magnetic fields and sense the induced B and H signals) and a control system (to apply in-phase 2D excitation signals and amplify, filter, and record the B and H data). Hysteresis loops were tested at low frequencies (0.1–9 Hz) and under different magnetization states (0.1–0.6 T) in two orientations—parallel and perpendicular to the annealing magnetic field—to verify anisotropy under varying conditions. Initial permeability, near-saturation magnetization, and basic magnetization curves (BM curves) were measured across different directions to provide parameters for simulations and theoretical calculations. Based on these measurements and finite element simulations, a mathematical model was developed to adjust the empirical coefficient λ used in theoretical SF calculations. The results revealed that the ratio of empirical coefficients in different directions is inversely proportional to the ratio of magnetic permeability in the corresponding directions. A verification group was established to compare the original model and the modified model. The mean squared error (MSE) between the original model and the finite element simulation was 49.97, while the MSE between the improved model and the finite element simulation was reduced to 0.13. This indicates a substantial improvement in the computational accuracy of the modified model.
With the breakthroughs in quantum theory and the rapid advancement of quantum precision measurement sensor technologies, atomic magnetometers based on the spin-exchange relaxation-free (SERF) mechanism have played an increasingly important role in ultra-weak biomagnetic field detection, inertial navigation, and fundamental physics research. To achieve high-precision measurements, SERF magnetometers must operate in an extremely weak magnetic field environment, while the detection of ultra-weak magnetic signals relies on a low-noise background. Therefore, accurate measurement, modeling, and analysis of magnetic noise in shielding materials are of critical importance. In this study, the magnetic noise of permalloy sheets was modeled, separated, and analyzed based on their measured magnetic properties, providing essential theoretical and experimental support for magnetic noise evaluation in shielding devices. First, a single-sheet tester (SST) was modeled via finite element analysis to investigate magnetization uniformity, and its structure was optimized by adding a supporting connection plate. Second, an experimental platform was established to verify magnetization uniformity and to perform accurate low-frequency measurements of hysteresis loops under different frequencies and field amplitudes while ensuring measurement precision. Finally, the Bertotti loss separation method combined with a PSO optimization algorithm was employed to accurately fit and analyze the three types of losses, thereby enabling precise separation and calculation of hysteresis loss. This provides essential theoretical foundations and primary data for magnetic noise evaluation in shielding devices.