The first water-cooling nano multilayer Kirkpatrick–Baez mirror system in the Structural Dynamics beamline (ID23) at High Energy Photon Source (HEPS) has been implemented. An Invar gantry is engineered to achieve a balance between light weight and stability. The cooling system, including eutectic gallium–indium (eGaIn), copper braids and a cooling water circuit with a multi-bend copper pipe in a compact space, decouples movement and mitigates vibration. The mirror cooling holders introduce less than 0.5 nm RMS height error and 0.1 µrad RMS slope error in the mirror surface shapes. A series of stability tests is applied to verify the mechanism structure. In a 1 h test assessment using a water-cooling flow of 4.5 L min −1 , the system demonstrated a positional stability of 5.96 nm and an angular stability of 86.76 nrad from 1 Hz to 500 Hz. The focal spot size of 13.39 nm × 15.15 nm (H × V) at a photon energy of 21.8 keV demonstrates the system's performance.
A metrology method dedicated to position monitoring and correction for the ptychography setup at the High Energy Photon Source (HEPS, Beijing, China) beamline ID09 is proposed. The sophisticated design of the Invar interferometer metrology frame ensures optimal performance within the compact environment between the focal spot and the sample setup. The measurement method and error analysis are described in detail. The sample setup achieves a positional stability of less than 2 nm RMS from 1 Hz to 500 Hz. The 12.3 nm reconstruction resolution of 2D ptychography at 12.4 keV demonstrates that the interferometer metrology strategy is available.