7 February 2026 in SSRN Electronic Journal
This article introduces a digitizer circuit for interfacing a four-wire configured resistive sensor (RX) with lead resistance compensation (RL) feature. The proposed circuit is minimalistic, utilizing only three core components: (i) an inverting amplifier, (ii) an integrator, and (iii) a comparator. The RX is integrated in the feedback path of the inverting amplifier which generates the required negative potentials for the dual-slope integrator to operate efficiently; highlighting circuit’s first novelty. The circuit operates in two distinct modes: a charging mode and a discharging mode using a couple of switches. These modes produce two separate equations involving RX and RL. By solving these equations, RX is determined independent of RL. This advancement, coupled with the circuit’s compact, and cost-effective design, underscores its second novelty. The circuit is fabricated on PCB, and has been experimentally tested for a Pt100 RTD (Resistance Temperature Detector) sensor, where the temperature changes from 0 °C to 200 °C, causing RX to vary from 100 Ω to 175.84 Ω (full span). The circuit operates on a single input supply and completes the measurement within a single charge-discharge cycle, highlighting its additional features. The sensor current (IXRMS) ranges from 5.87 mA to 6.86 mA, resulting in power dissipation between 3.44 mW and 8.27 mW across the full span. The measurement time varies from 86.21 ms to 62.39 ms, with an average error of ±0.32 %.