Key Laboratory of Ultra-Weak Magnetic Field Measurement Technology, Ministry of Education, School of Instrumentation and Optoelectronic Engineering, Beihang University, Beijing, China
Auto-generated profile, based on publicly available information
Zhongxiang Jing
Key Laboratory of Ultra-Weak Magnetic Field Measurement Technology, Ministry of Education, School of Instrumentation and Optoelectronic Engineering, Beihang University, Beijing, China
High-power devices used in hospitals generate harmonic magnetic field disturbances inside magnetic shielding cylinders (MSCs). These disturbances exhibit axial-gradient spatial distributions, which substantially degrade the imaging quality of magnetocardiography (MCG). To address this issue, a high-precision nonlinear spatial gradient disturbance suppression method is proposed for single-open MSCs. First, a nonlinear gradient coil design is developed to generate variable magnetic field gradients, precisely matching the nonlinear gradient distribution induced by external magnetic disturbances in the MSC. Subsequently, the entire system is modeled, and a composite controller based on proportional-integral repetitive control-linear extended state observer (PIRC-LESO) is designed to enhance the suppression of magnetic field disturbances with harmonic characteristics. Finally, an active magnetic field disturbance suppression system was implemented to evaluate the effectiveness of the nonlinear spatial gradient disturbance suppression method. Experimental results demonstrate that the proposed method reduces the peak-to-peak magnetic field disturbance from 42.0 to 6.9 pT and significantly enhances the signal-to-noise ratio of the MCG signals.
27 October 2025 in Mechanical Systems and Signal Processing