Anıl Doğan Bektaş

Anıl Doğan Bektaş

Auto-generated profile, based on publicly available information

Anıl Doğan Bektaş

Mem Bahcelievler Hosp, Clin Child Hlth & Dis, Istanbul, Turkiye

Research Focus & Metrics

Academic Career Length

7Years (2020-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
16 December 2020 in Scientific Reports
Verified
Journal Article
Open access
PDF file
MDMuhterem Duyu
Muhterem Duyu
ABAnıl Dogan Bektas
Anıl Dogan Bektas
ZKZeynep Karakaya
Zeynep Karakaya
MBMeral Bahar
Meral Bahar
AGAybuke Gunalp
Aybuke Gunalp
OBOzlem Bozkurt
Ozlem Bozkurt
The objective of this study was to evaluate a novel microstream method by comparison with PaCO2 and the more standard mainstream capnometer in intubated pediatric patients. We hypothesized that the novel microstream method would superior compared to the traditional mainstream method in predicting PaCO2. This was a prospective single-center comparative study. The study was carried out on 174 subjects with a total of 1338 values for each method. Data were collected prospectively from mainstream and microstream capnometer simultaneously and compared with PaCO2 results. Although both mainstream PetCO2 (mainPetCO2) and microstream PetCO2 (microPetCO2) were moderately correlated (r = 0.63 and r = 0.68, respectively) with PaCO2 values, mainPetCO2 was in better agreement with PaCO2 in all subjects (bias ± precision values of 3.8 ± 8.9 and 7.3 ± 8.2 mmHg, respectively). In those with severe pulmonary disease, the mainPetCO2 and microPetCO2 methods were highly correlated with PaCO2 (r = 0.80 and r = 0.81, respectively); however, the biases of both methods increased (14.8 ± 9.1 mmHg and 16.2 ± 9.0 mmHg, respectively). In cases with increased physiologic dead space ventilation, the agreement levels of mainPetCO2 and microPetCO2 methods became distorted (bias ± precision values of 20.9 ± 11.2 and 25.0 ± 11.8 mm Hg, respectively) even though mainPetCO2 and microPetCO2 were highly correlated (r = 0.78 and r = 0.78, respectively). It was found that the novel microstream capnometer method for PetCO2 measurements provided no superiority to the traditional mainstream method. Both capnometer methods may be useful in predicting the trend of PaCO2 due to significant correlations with the gold standard measurement in cases with severe pulmonary disease or increased physiological dead space –despite reduced accuracy.

9 September 2023 in Journal of Pediatric Infection
Verified
Journal Article
Closed access
ABAnıl Doğan Bektaş
Anıl Doğan Bektaş
EYEmine Olcay Yasa
Emine Olcay Yasa
ZHZafer Habip
Zafer Habip
EKEsra Koçoğlu
Esra Koçoğlu

9 September 2023 in Journal of Pediatric Infection
Verified
Journal Article
Closed access
ABAnıl Doğan Bektaş
Anıl Doğan Bektaş
EYEmine Olcay Yasa
Emine Olcay Yasa
ZHZafer Habip
Zafer Habip
EKEsra Koçoğlu
Esra Koçoğlu