To overcome the shortcomings of traditional Czerny-Turner (C-T) spectrometers—such as significant optical aberrations, low spatial resolution, and image quality degradation over time. This paper proposes an optical aberration correction design method for C-T imaging spectrometers based on an externally mounted transmissive Schmidt correction module. The method involves placing a Schmidt corrector plate at the front end of the telescope to correct the aberrations of the overall optical system. Accordingly, a spaceborne C-T imaging spectrometer optical system was designed with a 3.4° field of view, 60mm focal length, F-number(F#) of 5, and an operating spectral range of 400~700nm. The optical design software CODEV was used for ray tracing, optimization, and performance analysis. The results show that the addition of the external Schmidt corrector effectively corrects the system's aberrations. Across the full field of view, the modulation transfer function (MTF) remains above 0.7 at the Nyquist frequency of 38.46 lp/mm, approaching the diffraction limit and meeting all design requirements. These findings demonstrate the feasibility of the proposed optical design method, which offers a reliable technical approach for post-deployment maintenance, aberration correction, and service life extension of remote sensing instruments.