15 July 2019 in IEEE Access
Electrical impedance tomography (EIT) is a promising medical technique for monitoring brain injury through the reflection of electrical impedance changes in internal brain tissues. However, the contact impedances at the electrode-scalp interfaces can still notably affect the EIT accuracy. Hence, we propose a method for real-time monitoring the multi-channel contact impedances of cerebral EIT using a reference electrode attached close to the head vertex. This method allows to approximate inter-channel differences and changes over time of multi-channel contact impedances. The experimental results on 36 volunteers show that the maximum relative inter-channel difference obtained from 16 disposable surface electrodes reaches 18.7%, and the maximum relative change over time reaches 19.2%. The proposed method can assist clinicians in real-time monitoring of contact conditions from the multiple electrode-scalp interfaces during the clinical application of the cerebral EIT.
1 January 2020 in Physiological Measurement
Objective: This study investigated the feasibility of electrical impedance tomography (EIT) for monitoring the deterioration of ischemic lesion after the onset of stroke. Approach: Fifteen rats were randomly distributed into two groups: rats operated to establish a right middle cerebral artery occlusion (MCAO) (n = 10), and sham-operated rats (n = 5). Then, the operated rats were kept 2 h under anesthesia for EIT monitoring. Subsequently, descriptive statistical analysis was performed on whole-brain resistivity changes, and repeated-measures analysis of variance (ANOVA) on the average resistivity variation index. Additionally, pathological examinations were performed after 6 h of infarction. Main results: The results obtained showed that ischemic damage developed in the right corpus striatum of the rats with MCAO, whereas the brains of the sham group showed no anomalies. The descriptive statistical analysis revealed that the whole-brain resistivity changes after 30, 60, 90, and 120 min of infarction were 0.063 ± 0.038, 0.097 ± 0.046, 0.141 ± 0.062, and 0.204 ± 0.092 for the rats with MCAO and 0.029 ± 0.021, 0.002 ± 0.002, 0.017 ± 0.011, and −0.001 ± 0.011 for the sham-operated rats, respectively. The repeated-measures ANOVA revealed that the right MCAO model resulted in a significant impedance increase in the right hemisphere, which continued to increase over time after infarction. Significance: The overall study results indicate that EIT facilitates monitoring of local impedance variations caused by MCAO and may be a solution for real-time monitoring of intracranial pathological changes in ischemic stroke patients.
29 July 2019 in BioMedical Engineering OnLine
Electrical impedance tomography (EIT) is a noninvasive, radiation-free, and low-cost imaging modality for monitoring the conductivity distribution inside a patient. Nowadays, time-difference EIT (tdEIT) is used extensively as it has fast imaging speed and can reflect the dynamic changes of diseases, which make it attractive for a number of medical applications. Moreover, modeling errors are compensated to some extent by subtraction of voltage measurements collected before and after the change. However, tissue conductivity varies with frequency and tdEIT does not efficiently exploit multi-frequency information as it only uses measurements associated with a single frequency. This paper proposes a tdEIT algorithm that imposes spectral constraints on the framework of the linear least squares problem. Simulation and phantom experiments are conducted to compare the proposed spectral constraints algorithm (SC) with the damped least squares algorithm (DLS), which is a stable tdEIT algorithm used in clinical practice. The condition number and rank of the matrices needing inverses are analyzed, and image quality is evaluated using four indexes. The possibility of multi-tissue imaging and the influence of spectral errors are also explored. Significant performance improvement is achieved by combining multi-frequency and time-difference information. The simulation results show that, in one-step iteration, both algorithms have the same condition number and rank, but SC effectively reduces image noise by 20.25% compared to DLS. In addition, deformation error and position error are reduced by 8.37% and 7.86%, respectively. In two-step iteration, the rank of SC is greatly increased, which suggests that more information is employed in image reconstruction. Image noise is further reduced by an average of 32.58%, and deformation error and position error are also reduced by 20.20% and 31.36%, respectively. The phantom results also indicate that SC has stronger noise suppression and target identification abilities, and this advantage is more obvious with iteration. The results of multi-tissue imaging show that SC has the unique advantage of automatically extracting a single tissue to image. SC enables tdEIT to utilize multi-frequency information in cases where the spectral constraints are known and then provides higher quality images for applications.
27 June 2019 in Medical & Biological Engineering & Computing
27 September 2019 in IEEE Access
The objective of the study was to develop an optimized method for dynamic electrical impedance tomography (EIT) to image large area conductive perturbation (LACP), a new type of imaging target that we found during the monitoring and evaluation of the mannitol dehydration treatment of patients with brain edema based on brain EIT. Previously, we reconstructed LACP images with the commonly used NOSER algorithm with polar driven pattern. However, conductivity changes near the center of the LACP were blurred or remained undetected, and blocky artifacts appeared in the reconstructed images, making the interpretation of the results difficult. To solve this problem, we, for the first time, propose an optimized algorithm for imaging LACP. This algorithm comprises a modified sensitivity matrix to compensate for the blurred conductivity changes near the center. It also uses the Markov random field constraint to reduce blocky artifacts. To verify the performance of the proposed method, we conducted experiments based on head models and human subjects. Specific metrics, including shape error (SE) and image fluctuation (IF) artifacts, were also proposed to evaluate the image quality of the LACP. The experimental results demonstrated that, compared with the NOSER algorithm, the proposed method could respectively reduce SE and IF by 51.7% and 47.5%. Therefore, it can optimize the imaging of LACP and provide references for other applications at present or in the future which involve imaging LACP.
1 March 2020 in Physiological Measurement
Objective: Electrode detachment may occur during dynamic brain electrical impedance tomography (EIT) measurements. After the faulty electrodes have been reset, EIT can restore to steady monitoring but the corrupted data, which will challenge interpretation of the results, are notoriously difficult to recover. Approach: Here, a piecewise processing method (PPM) is introduced to manage the erroneous EIT data after reattachment of faulty electrodes. In the PPM, we define the three phases before, during and after reconnection of the faulty electrode as PI, PII and PIII, respectively. Using this definition, an empirical mode decomposition-based interpolation method is introduced to compensate the corrupted data in PII, using the valid measurements in PI and PIII. Then, the compensated data in PII are spliced at the end of PI. Thus, there will be a surge at the junction of PII and PIII due to the changes in contact state of the repositioned electrodes. Finally, to ensure all the EIT data are obtained under constant electrode settings, we calculate the above changes and eliminate them from the data after PII. To verify the performance of the PPM, experiments based on head models, with anatomical structures and with human subjects were conducted. Metrics including permutation entropy (PE) and image correlation (IC) were proposed to measure the stability of the signal and the quality of the reconstructed EIT images, respectively. Main results: The results demonstrated that the PE of the processed data was reduced to 0.25 and the IC improved to 0.78. Significance: Without iterative calculations the PPM could efficiently manage the erroneous EIT data after reattachment of the faulty electrodes.