1 October 2015 in Chinese Physics C
The resistance to neutron irradiation of LaBr3 scintillator was studied in this work. The change of background counting rate, light output and energy resolution of the LaBr3 scintillator were analyzed to determine whether the scintillator was damaged under different neutron flux rates induced by 241Am-Be, D-T neutron generator, and reactor neutron source. The results show that the neutron radiation damage in LaBr3 scintillator is mainly affected by neutron flux rate. Under low flux rate, the properties of the scintillator were hardly changing, while under high flux rate, there is obvious deterioration in the background spectra and in the energy resolution because of the neutron activation. After a period, the neutron radiation damage will spontaneously recover.
20 June 2023 in SCIENTIA SINICA Physica, Mechanica & Astronomica
A major method to obtain the characteristic information about equipment with transient pulse radiation is to detect the gamma rays released from the equipment. To explore the performances of the equipment, sophisticated diagnostic techniques and detecting systems have been developed and established. The detecting system serves as the main instrumentation for gathering details about radiation fields. It also serves as the technological basis for carrying out intense pulsed radiation detection, and its performance directly influences the selection of the measurement technique and the quality of the measured data. The detection principle, detector technology, and development situation of intense pulsed gamma rays are discussed in this study.
10 February 2022 in IEEE Transactions on Nuclear Science
The energy thresholding characteristic of carbon dioxide (CO 2 ) Cherenkov gas is significant for the detection of fusion $\gamma $ -rays. In this article, the energy thresholding characteristic of CO 2 (at 300 K and 4.02 atm.) gas is experimentally calibrated by a pulse electron beam with picosecond duration time and adjustable energies. Three mirrors with the ultraviolet reflectivity are used to make an off-axis light collection system. For electron energy of 10 MeV (above the Cherenkov threshold energy), the collected pulse photon spectrum has good agreement with the theoretical spectrum of Cherenkov radiation, confirming that Cherenkov radiation dominates collected photons than non-Cherenkov sources. The threshold energy resolution (12/7.5 MeV) is improved to ~10 4 , an order higher than the reported value of ~10 3 . The improvement on the threshold energy resolution of CO 2 gas is mainly attributed to the ultraviolet mirrors, suppressing lights with a wavelength higher than 280 nm. This work can provide significant support for the further application of the Cherenkov detector in the fusion $\gamma $ -ray diagnosis.