<!-- *** Custom HTML *** --> Purpose: The impact of system parameters on signal detectability can be studied with simulation platforms. We describe the steps taken to verify and confirm the accuracy of a local platform developed for the use in virtual clinical trials. Approach: The platform simulates specific targets into existing two-dimensional full-field digital mammography and digital breast tomosynthesis images acquired on a Siemens Inspiration system. There are three steps: (1) creation of voxel models or analytical objects; (2) generation of a realistic object template with accurate resolution, scatter, and noise properties; and (3) insertion and reconstruction. Four objects were simulated: a 0.5-mm aluminium (Al) sphere and a 0.2-mm-thick Al sheet in a PMMA stack, a 0.8-mm steel edge and a three-dimensional mass model in a structured background phantom. Simulated results were compared to acquired data. Results: Peak contrast and signal difference-to-noise ratio (SDNR) were in close agreement (<5 % error) for both sphere and sheet. The similarity of pixel value profiles for sphere and sheet in the xy direction and the artifact spread function for real and simulated spheres confirmed accurate geometric modeling. Absolute and relative average deviation between modulation transfer function measured from a real and simulated edges showed accurate sharpness modelling for spatial frequencies up to the Nyquist frequency. Real and simulated objects could not be differentiated visually. Conclusions: The results indicate that this simulation framework is a strong candidate for use in virtual clinical studies.
<!-- *** Custom HTML *** --> Purpose: The relevance of presampling modulation transfer function (MTF) measurements in digital mammography (DM) quality control (QC) is examined. Two studies are presented: a case study on the impact of a reduction in MTF on the technical image quality score and analysis of the robustness of routine QC MTF measurements. Approach: In the first study, two needle computed radiography (CR) plates with identical sensitivities were used with differences in the 50% point of the MTF (fMTF0.5) larger than the limiting value in the European guidelines (>10 % change between successive measurements). Technical image quality was assessed via threshold gold thickness of the CDMAM phantom and threshold microcalcification diameter of the L1 structured phantom. For the second study, presampling MTF results from 595 half-yearly QC tests of 55 DM systems (16 types, six manufacturers) were analyzed for changes from the baseline value and changes in fMTF0.5 between successive tests. Results: A reduction of 20% in fMTF0.5 of the two CR plates was observed. There was a tendency to a lower score for task-based metrics, but none were significant. Averaging over 55 systems, the absolute relative change in fMTF0.5 between consecutive tests (with 95% confidence interval) was 3% (2.5% to 3.4%). Analysis of the maximum relative change from baseline revealed changes of up to −10 % for one a-Se based system and −15 % for a group of CsI-based systems. Conclusions: A limit of 10% is a relevant action level for investigation. If exceeded, then the impact on performance has to be verified with extra metrics.