We developed a multi-beam X-ray imaging detector, consisting of four scintillator screens connected by a branched optical fiber bundle with a CMOS camera. By using the detector and a multi-beam imaging optics with silicon single crystalline blades designed for a white synchrotron radiation source, we successfully demonstrated multi-beam X-ray imaging with an exposure time of 1 ms. The long and flexible optical fiber bundle used for the detector enables us to realize high-speed multi-beam X-ray imaging with high flexibility at a relatively low cost.
29 July 2019 in Japanese Journal of Applied Physics
We propose a novel approach to fabricate hard-X-ray optics using an ultracentrifuge machine. As a demonstration, we fabricated an X-ray absorption grating made of gold with a pitch of 30 μ m and an effective depth of 44 μ m in 67 μ m depth grooves [a filling rate of 65%, as same as the filling rate of random packing for equal-sized spheres in a jammed state (64%)]. Our approach is cost-efficient and has a high production-rate and material-selectivity, and will be widely used for fabricating optics not only for X-rays but also neutrons in the near future.
26 October 2017 in The Journal of Physical Chemistry C
An X-ray reflectometer using a laboratory X-ray source for quick measurements of the specular X-ray reflectivity curve is presented. It uses a bent–twisted crystal to monochromatize and focus the diverging X-rays (Cu Kα1) from a laboratory point source onto the sample. The reflected X-rays are recorded with a two-dimensional detector. Reflectivity curves can be measured without rotating the sample, detector or X-ray source during measurements. The instrument can separate the specularly reflected X-rays from the diffuse scattering background, so low reflectivities can be measured accurately. For a gold thin film on silicon, the reflectivity down to the order of 10−6 was obtained with a measurement time of 100 s and that down to 10−5 with a measurement time of 10 s. Reflectivity curves of a silicon wafer and a liquid ethylene glycol surface are shown as well. Time-resolved measurements of a TiO2 surface during UV irradiation are also reported.