8 July 2020 in IEEE Transactions on Circuits and Systems II: Express Briefs
Achieving a high input impedance ${Z} _{\mathrm{ in}}$ is important for the instrumentation amplifier (IA) performing neural and biopotential signal sensing. Previously, impedance boosting techniques have been used with a positive feedback loop (PFL) driven by the output or auxiliary (AUX) path driven by the input. We propose that the conventional approach to DC-servo loop (DSL), which blocks the electrode offset (EOS), does not fully enable the function of a PFL or AUX path. This results in either limited impedance boosting or reduction of a low-frequency ${Z} _{\mathrm{ in}}$ in the sub-Hz band. To solve this problem and achieve ${Z} _{\mathrm{ in }}$ boosting down to sub-Hz, we propose a new approach to DSL for a capacitively-coupled chopper instrumentation amplifier (CCIA). The proposed DSL is realized using a 30-nA opamp, and the AUX path is designed using a 450-nA duty-cycled buffer. The CCIA achieves a gain of 40 dB and is implemented using a $0.18~\mu \text{m}$ CMOS process with an area of 0.19 mm 2 . The measured results show that ${Z} _{\mathrm{ in}}$ is increased by 184 times up to 4.6 $\text{G}{\Omega }$ at 0.01 Hz. Compared to the CCIA using conventional DSL, ${Z} _{\mathrm{ in}}$ is improved by 7.8 times. An input-referred noise of $2.1~\mu \text{V}_{\mathrm{ rms}}$ is achieved at $2.14~\mu \text{W}$ .