13 July 2020 in Metrologia
After several years of development and continuous improvement of the BIPM Kibble balance, mass measurement of a 1 kg Pt/Ir artefact in terms of the Planck constant has been carried out. The paper describes the apparatus and its alignment, and explains the measurement and data processing. The balance operates in a one-mode and two measurement phase scheme using a bifilar coil. The mass value of the artefact was measured in vacuum with a relative standard uncertainty of 49 µg. This value differs from the value deduced from the calibration following the dissemination process after the redefinition of the kilogram by 15 µg.
28 December 2017 in Metrologia
The Kibble balance is expected to become an important instrument in the near future for realizing the unit of mass, the kilogram, in the revised international system of units (SI). The Kibble balance assumes an equality of two magnetic profiles measured in the weighing and velocity phases. A recent study conducted in the Kibble balance group at the Bureau International des Poids et Mesures (BIPM) showed that the coil current could significantly affect the magnetic profile, which should be carefully taken into account in the Kibble balance experiment. This paper gives a deeper understanding and investigation of the effect, and discusses the magnetic profile change due to the coil current, for both the classical two-mode and the one-mode Kibble balances. The coil current effect has been theoretically and experimentally investigated based on a typical magnet design with an air gap. One important conclusion found in the one-mode Kibble balance is that the magnetic profile change measured in the velocity phase is twice the change in the weighing phase. A compensation suggestion, to minimize the profile change due to the coil current in a BIPM-type magnet, is presented.
11 September 2017 in Metrologia
The magnet is one of the fundamental elements in Kibble balances. The Kibble balance group at the Bureau International des Poids et Mesures (BIPM) proposed a permanent magnet with a radial field, which has been widely employed in other Kibble balance experiments. In this paper, we discuss a different method of operation for the BIPM-type magnet, i.e. only the lower half of the magnetic circuit is closed. The merit of such a magnet is its convenience for opening the air gap and adjusting the coil. A disadvantage is that it can yield a lower and sloped magnetic field profile with less shielding. In the approach described, high permeability yokes are used to flatten the magnetic profile, which has proven to be a novel and convenient approach for removing the nonuniformity of the magnetic profile due to magnet magnetization asymmetries. The performance of the half-closed BIPM magnet is evaluated using experimental measurements. The results show that the half-closed magnetic circuit retains most of the main features of the fully-closed magnet and is a good option for the early stages of Kibble balance operations. In addition, the profile change due to the coil current is predicted and experimentally verified with a bifilar coil. Related systematic effects are discussed for the Planck constant measurement.
1 January 2023 in Metrologia
Main text: This report describes the second CCM key comparison of realizations of the kilogram definition based on the fixed numerical value of the Planck constant, which came into force on 20 May 2019. The objectives were to determine the level of agreement between realizations of the kilogram using Kibble and joule balances and the x-ray crystal density (XRCD) method and to provide input for the calculation of the second "consensus value" of the kilogram. The consensus value serves as the basis for an internationally coordinated dissemination of the kilogram and is updated after each new key comparison. Its use will continue until satisfactory agreement between realization experiments has been achieved. Another objective was the determination of the reproducibility of the realization experiments by comparing the new results with those of the first key comparison of kilogram realizations, CCM.M-K8.2019. The comparison was organized by the BIPM and had nine participants. The BIPM, LNE, METAS, NIST, NRC and UME operated Kibble balances, the NIM used a joule balance and the NMIJ and the PTB participated using 28Si spheres, the masses of which were determined with the XRCD method. These realization methods were used to calibrate 1 kg mass standards under vacuum or in air. The standards were sent to the BIPM where they were compared with each other and with BIPM Pt-Ir working standards. The latter were calibrated traceable to the International Prototype of the Kilogram (IPK), the mass of which served as the definition of the kilogram until 20 May 2019. The results of the weighings at the BIPM together with the measurement results communicated by the participants allowed comparison of the values attributed to 1 kg mass standards using the realization experiments of the participants. The level of agreement between mass determinations with the realization experiments and the BIPM as-maintained mass unit, traceable to the Planck constant through the mass of the International Prototype of the Kilogram, could also be deduced. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
12 October 2015 in Metrologia
The International Bureau of Weights and Measures (BIPM) is developing a watt balance for the forthcoming redefinition of the kilogram. An improved version of the apparatus, based on a new closed magnetic circuit is now being assembled. The new apparatus will significantly reduce the type B uncertainty due to misalignment of the magnetic circuit as this work demonstrates. We present two techniques recently developed to accurately align the magnetic field of the circuit perpendicular to the direction defined by the local acceleration of gravity. Uncertainty below 30 μrad was achieved for both techniques which fulfils the requirements for the BIPM watt balance to enable a Planck constant determination at the 1 × 10−8 level.
30 December 2019 in IEEE Transactions on Instrumentation and Measurement
Yoke-based permanent magnetic circuits are widely used in Kibble balance experiments. In these magnetic systems, the coil current, with positive and negative signs in two steps of the weighing measurement, can cause an additional magnetic flux in the circuit and hence a magnetic field change at the coil position. The magnetic field change due to the coil current and related systematic effects have been studied with the assumption that the yoke material does not contain any magnetic hysteresis. In this paper, we present an explanation of the magnetic hysteresis error in Kibble balance measurements. An evaluation technique based on measuring yoke minor hysteresis loops is proposed to estimate the effect. The dependence of the magnetic hysteresis effect and some possible optimizations for suppressing this effect are discussed.