The modular inner branched stent-graft (MIBSG), a novel interventional therapy, has demonstrated good effects in the endovascular treatment of aortic arch pathologies, especially those involving the supra-aortic branches. Nevertheless, the long-term efficacy of the MIBSG and in-depth quantitative evaluation of postoperative outcomes remain to be examined. Moreover, the regularity of postoperative vascular remodeling induced by MIBSG implantation has yet to be explored. To address these questions, we constructed four models (normal, preoperative, 1 week postoperative, and 6 months postoperative) based on a single patient case to perform computational fluid dynamics simulations. The morphological and hemodynamic characteristics, including the velocity profile, flow rate distribution, and hemodynamic parameter distribution (wall shear stress and its derivative parameters), were investigated. After MIBSG implantation, the morphology of the supra-aortic branches changed significantly, and the branch point moved forward to the proximal ascending aorta. Moreover, the curvature radius of the aortic arch axis continued to change. These changes in morphology altered the characteristics of the flow field and wall shear stress distribution. As a result, the local forces exerted on the vessel wall by the blood led to vessel remodeling. This study provides insight into the vascular remodeling process after MIBSG implantation, which occurs as a result of the interplay between vascular morphological characteristics and blood flow characteristics.
Objectives: The aim of this study was to evaluate the feasibility and safety of a non-customized modular inner branched stent graft for total endovascular aortic arch repair in a porcine model. Methods: The modular inner branched stent graft system with a split main body design included 1 proximal main component, 1 distal main component, and 1 branched covered stent. The gutter in the proximal main component was sealed with sutured membrane. Fatigue testing was performed to evaluate the durability of the stent graft. Fifteen pigs were used in this study. In each pig, a stent graft was delivered and deployed to the aortic arch through the femoral arterial access and right carotid arterial access. Angiography and computed tomography angiography were used to evaluate the morphological features before euthanasia. After euthanasia, the implanted device, surrounding tissue, and major organs were harvested for gross and histological examination. Results: There were no collapses and no stent graft fractures detected after fatigue testing. The technical success rate was 14/15, and the incidence of major adverse cardiovascular events was 2/15. Angiography performed at the end of follow-up revealed no endoleaks and no device migration. Histological examination demonstrated excellent biocompatibility of the stent graft. Conclusions: The non-customized modular inner branched stent graft system is safe and feasible for the endovascular reconstruction of the aortic arch in a porcine model.