Elisabeth Debout

Ms. Elisabeth Debout


Research Focus & Metrics

Academic Career Length

7Years (2020-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
24 April 2020 in European Spine Journal
Verified
Journal Article
Open access
PDF file
LZLaura Zengerle
Laura Zengerle
AKAnne Köhler
Anne Köhler
EDElisabeth Debout
Elisabeth Debout
CHCarsten Hackenbroch
Carsten Hackenbroch
HWHans-Joachim Wilke
Hans-Joachim Wilke
Purpose: Disc herniations are usually treated by decompression of the spinal nerves via a partial nucleotomy. As a consequence of reduced disc height (DH), reduced intradiscal pressure (IDP) and increased range of motion (ROM), accelerated degeneration may occur. Nucleus replacement implants are intended to restore those values, but are associated with the risk of extrusion. Methods: In six fresh frozen lumbar spinal segments (L2-3/L3-4/L4-5/L5-S1, age median 64.5 years (57–72), Pfirrmann grade 2–3), a prolapse was provoked through a box defect (6 × 10 mm) in the annulus. The herniated nucleus material was removed and replaced by a novel collagen-based nucleus implant. An annulus closure device sealed the defect. ROM, neutral zone (NZ) and IDP were measured in the (1) intact and (2) defect state, (3) postoperatively and (4) after cyclic loading (n = 100,000 cycles) applying pure moments (± 7.5 Nm) in flexion–extension, lateral bending and axial rotation. Additionally, the change in DH was determined. Extrusion of implants or nucleus material was evaluated macroscopically. Results: In all specimens, a prolapse could be provoked which decreased DH. Subsequent nucleotomy changed ROM/NZ and IDP considerably. Initial values could be restored by the implantation. Macroscopically, none of the implants nor nucleus material did migrate after cyclic loading. Conclusions: In this study, a prolapse followed by a nucleotomy resulted in a biomechanical destabilisation. Implantation of the nucleus replacement combined with an annulus closure restored the intact condition without showing signs of extrusion nor migration after cyclic loading. Hence, nucleus replacements could have a new chance in combination with annulus closure devices.

22 March 2021 in Applied Sciences
Verified
Journal Article
Open access
PDF file
LZLaura Zengerle
Laura Zengerle
EDElisabeth Debout
Elisabeth Debout
BKBruno Kluger
Bruno Kluger
LZLena Zöllner
Lena Zöllner
HWHans-Joachim Wilke
Hans-Joachim Wilke
Lumbar disc herniation (LDH) is the most common reason for low back pain in the working society. New regenerative approaches and novel implants are directed towards the restoration of the disc or its biomechanical properties. Aiming to investigate these new therapies under physiological conditions, in this study, a model for LDH was established by developing a new physiological in vitro test method. In 14 human lumbar motion segments, different daily-life and worst-case activities were simulated successfully by applying a physiological range of motion and axial loading in order to create physiological intradiscal pressure. An LDH could be provoked in 11 of the 14 specimens through vertical and round annular defects of different sizes. Interestingly, the defect and the LDH hardly influenced the biomechanical properties of the disc. For the investigation of regenerative approaches in further experiments, the recommendation based on the results of this study is to create an LDH in non-degenerated motion segments by the application of the new physiological in vitro test method after setting the round annular defects to a size of 4 mm in diameter.

27 May 2020 in European Spine Journal
Verified
Journal Article
Open access
PDF file
LZLaura Zengerle
Laura Zengerle
AKAnne Köhler
Anne Köhler
EDElisabeth Debout
Elisabeth Debout
CHCarsten Hackenbroch
Carsten Hackenbroch
HWHans-Joachim Wilke
Hans-Joachim Wilke
Unfortunately, Fig. 7 and last paragraph of the result section have been incorrectly published. The complete corrected Fig. 7 and last paragraph