DMS Technology is based on ion segregation phenomenon when flowing through the volume of the detector. DMS detector is made up with two parallel ceramic plates with electrodes layered on each plate. High voltage and high frequency of electric field generated in detector’s volume causes ion segregation on collection electrode (see Fig. 1).
Fig. 1Diagram of DMS detector composition.
Observed ion segregation in flowing gas is caused by different mobilities of the analyzed chemicals in fields characterized by lower and higher intensity. Ion mobility value is dependent on ion mass, charge and velocity of flowing gas [1–4]. Figure 2 depicts example flow gas through detector’s volume. Variable electric field put on electrodes causes capturing of these ions, which mobility does not enable stable (parallel to detector’s plates) flow through detector’s volume. Thus, phenomenon of ion mobility of ions flowing through the detector variable on compensated electrical field, makes detector a