To measure the absorbed dose rate to water and penumbra of a Gamma Knife® (GK) using a polymethyl metacrylate (PMMA) phantom. A multi-purpose PMMA phantom was developed to measure the absorbed dose rate to water and the dose distribution of a GK. The phantom consists of a hemispherical outer phantom, one exchangeable cylindrical chamber-hosting inner phantom, and two film-hosting inner phantoms. The radius of the phantom was determined considering the electron density of the PMMA such that it corresponds to 8 g/cm2 water depth, which is the reference depth of the absorbed dose measurement of GK. The absorbed dose rate to water was measured with a PTW TN31010 chamber, and the dose distributions were measured with radiochromic films at the calibration center of a patient positioning system of a GK Perfexion. A spherical water-filled phantom with the same water equivalent depth was constructed as a reference phantom. The dose rate to water and dose distributions at the center of a circular field delimited by a 16-mm collimator were measured with the PMMA phantom at six GK Perfexion sites. The radius of the PMMA phantom was determined to be 6.93 cm, corresponding to equivalent water depth of 8 g/cm2. The absorbed dose rate to water was measured with the PMMA phantom, the spherical water-filled phantom and a commercial solid water phantom. The measured dose rate with the PMMA phantom was 1.2% and 1.8% higher than those measured with the spherical water-filled phantom and the solid water phantom, respectively. These differences can be explained by the scattered photon contribution of PMMA off incoming 60Co gamma-rays to the dose rate. The average full width half maximum and penumbra values measured with the PMMA phantom showed reasonable agreement with two calculated values, one at the center of the PMMA phantom (LGP6.93) and other at the center of a water sphere with a radius of 8 cm (LGP8.0) given by Leksell Gamma Plan using the TMR10 algorithm. A PMMA phantom constructed in this study to measure the absorbed dose rates to water and dose distributions of a GK represents an acceptable and practical alternative for GK dosimetry considering its cost-effectiveness and ease of handling.
Three-dimensional printing technology has the advantage of facilitating the construction of complex three-dimensional shapes. For this reason, it is widely used in medical and radiological fields. However, few materials with high electron density similar to that of bone exist for fabricating a human phantom. In this study, commercially available filament materials were used with an FDM 3D printer to perform delivery quality assurance (DQA) and were evaluated for medical use. For the bone filament material, BaSO4 was synthesized in five ratios of 2%, 4%, 6%, 8%, and 10% with 40% PBAT and 50~58% PLA. The electron density for the 3D printing material fabricated was obtained using kV energy CT and compared with the electron density of human organs and bones. The radiation beam properties of the 3D printed structures were analyzed as films for treatment using a linear accelerator. As a result, by changing the infill density of the material, it was possible to produce a material similar to the density of human organs, and a homogeneous bone material with HU values ranging from 371 ± 9 to 1013 ± 28 was produced. The 3D printing material developed in this study is expected to be usefully applied to the development of a patient-specific phantom to evaluate the accuracy of radiotherapy.
In Korea, when replacing or discarding parts of a medical linear accelerator (linac), self-disposal is required in the consideration of the activity, but there is no standard regulation to manage radioactive waste. The aim of this study is to check the activity of each part to determine the disposal time according to the clearance level for self-disposal. The results of measuring the components of the linac head parts of the disposed Varian, Elekta, and Siemens equipment were reflected in the Monte Carlo simulation to confirm the radionuclide change according to the presence or absence of impurities. To confirm the degree of activation of the linac, the main radionuclides according to the time after the linac shutdown, considering the workloads of 40/80 Gy/day of 10/15 MV linac irradiated with beams for 10 years in the results of the simulation of the linac parts, and the radionuclide concentration was confirmed. As a result of applying the clearance level for self-disposal in the notice of the Korean Nuclear Safety (KINS) to each linac head part, most parts of the 10 MV linac could be dismantled after 1 month, and 15 MV target and primary collimators were stored after a long period of time before being dismantled. Although additional radionuclides were identified according to the presence or absence of impurities, the disposal timing for each part did not change significantly. In this study, the clearance level for self-disposal for each radionuclide was applied to activated parts by three manufacturers to confirm the self-disposal timing and predict the timing at which workers are not exposed to radiation during dismantling/disposal.
19 June 2023 in Nuclear Engineering and Technology