AGNES (absence of gradients and Nernstian equilibrium stripping) is a stripping technique consisting of two conceptual steps: (i) application of a potential program (e.g. a step at a fixed potential) generating a known concentration gain between the outer and inner concentrations of the metal at the mercury electrode surface together with null gradients of the concentration profiles (inside and outside the mercury electrode) and (ii) determination of the concentration of reduced metal inside the amalgam in a stripping step. In the present implementation, the stripping step under diffusion limited conditions leads to a measured current just proportional to the free metal ion concentration. In this paper we present the basic principles of the technique, analytical expressions for a simplified model of its voltammetric implementation and a numerical study for a more refined model together with preliminary experimental results in the Cd(II)/nitrilotriacetic acid system showing how this technique can be used as an alternative to other techniques (such as ion selective electrodes) in order to determine free metal activities or concentrations in the presence of complex mixtures avoiding complications such as electrodic adsorption or complexation kinetics
Absence of Gradients and Nernstian Equilibrium Stripping (AGNES) is a recently suggested electroanalytical technique designed for the determination of the free concentration of heavy metals (such as Zn, Cd or Pb) which is here developed and applied to seawater samples. A key improvement for the implementation of AGNES with complex matrices is the development of a new blank, called the shifted blank (presented in this work for the first time), which can be applied to the same sample where the measurement is intended. The careful selection of the required parameters for the determination of the free Zn concentration (or activity) at the nanomolar level is described in detail. The methodology has been validated with a synthetic solution containing Zn and nitrilotriacetic acid (NTA) and then applied, as a first case, to two coastal seawater samples taken close to Barcelona and Tarragona (Catalonia, North-Eastern Spain) finding values in the range of 1-3nM, representing around 25% of total Zn. This technique can, in the near future, be crucial in helping to elucidate the role of the free zinc(II) concentration in natural waters.
Absence of gradient and Nernstian equilibrium stripping (AGNES) senses the free ion concentration of Zn(II) in solutions containing different ligands, being unaffected by the lack of reversibility of the Zn2+/Zn0 couple under the conditions assayed. In the presence of oxalate, the determination of [Zn2+] agrees with the stability and solubility constants of this sparingly soluble salt, once the precipitation kinetics are taken into account. Different strategies have been analysed and implemented in order to reduce the preconcentration time with the standard electrode of the polarographic stand (smallest drop radius around 0.141 mm): (i) using a lower preconcentration factor when there is no need of enhanced limit of detection; (ii) splitting the deposition stage into two, with a first potential step under diffusion limited conditions; (iii) the analysis of the chronoamperometric response in the deposition stage allows its duration to be adjusted, especially if non-inert complexes contribute to the arriving flux of metal to the mercury electrode. The two-potential-steps strategy is assessed as the most suitable in a general case
The determination of free Zn(2+) ion concentration is a key in the study of environmental systems like river water and soils, due to its impact on bioavailability and toxicity. AGNES (Absence of Gradients and Nernstian Equilibrium Stripping) and DMT (Donnan Membrane Technique) are emerging techniques suited for the determination of free heavy metal concentrations, especially in the case of Zn(2+), given that there is no commercial Ion Selective Electrode. In this work, both techniques have been applied to synthetic samples (containing Zn and NTA) and natural samples (Rhine river water and soils), showing good agreement. pH fluctuations in DMT and N(2)/CO(2) purging system used in AGNES did not affect considerably the measurements done in Rhine river water and soil samples. Results of DMT in situ of Rhine river water are comparable to those of AGNES in the lab. The comparison of this work provides a cross-validation for both techniques.