Siyu Yang

Dr. Siyu Yang

orcid

Research Focus & Metrics

Academic Career Length

6Years (2021-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
Verified
Journal Article
Open access
PDF file
SYSiyu Yang
Siyu Yang
XXXian Xian
Xian Xian
YLYiteng Liu
Yiteng Liu
ZZZiyi Zhang
Ziyi Zhang
QSQi Song
Qi Song
WWWeijia Wen
Weijia Wen
Polydimethylsiloxane (PDMS) has been widely used to make lab-on-a-chip devices, such as reactors and sensors, for biological research. Real-time nucleic acid testing is one of the main applications of PDMS microfluidic chips due to their high biocompatibility and transparency. However, the inherent hydrophobicity and excessive gas permeability of PDMS hinder its applications in many fields. This study developed a silicon-based polydimethylsiloxane-polyethylene-glycol (PDMS-PEG) copolymer microfluidic chip, the PDMS-PEG copolymer silicon chip (PPc-Si chip), for biomolecular diagnosis. By adjusting the modifier formula for PDMS, the hydrophilic switch occurred within 15 s after contact with water, resulting in only a 0.8% reduction in transmittance after modification. In addition, we evaluated the transmittance at a wide range of wavelengths from 200 nm to 1000 nm to provide a reference for its optical property study and application in optical-related devices. The improved hydrophilicity was achieved by introducing a large number of hydroxyl groups, which also resulted in excellent bonding strength of PPc-Si chips. The bonding condition was easy to achieve and time-saving. Real-time PCR tests were successfully conducted with higher efficiency and lower non-specific absorption. This chip has a high potential for a wide range of applications in point-of-care tests (POCT) and rapid disease diagnosis.

4 May 2021 in ACS Applied Bio Materials
Verified
Journal Article
Closed access
SYSiyu Yang
Siyu Yang
WWWeijia Wen
Weijia Wen

21 January 2023 in Molecules
Verified
Journal Article
Open access
PDF file
SYSiyu Yang
Siyu Yang
ZZZiyi Zhang
Ziyi Zhang
XXXian Xian
Xian Xian
QSQi Song
Qi Song
YLYiteng Liu
Yiteng Liu
WWWeijia Wen
Weijia Wen
Real-time polymerase chain reaction (real-time PCR) tests were successfully conducted in an aluminum-based microfluidic chip developed in this work. The reaction chamber was coated with silicone-modified epoxy resin to isolate the reaction system from metal surfaces, preventing the metal ions from interfering with the reaction process. The patterned aluminum substrate was bonded with a hydroxylated glass mask using silicone sealant at room temperature. The effect of thermal expansion was counteracted by the elasticity of cured silicone. With the heating process closely monitored, real-time PCR testing in reaction chambers proceeded smoothly, and the results show similar quantification cycle values to those of traditional test sets. Scanning electron microscope (SEM) and atomic force microscopy (AFM) images showed that the surface of the reaction chamber was smoothly coated, illustrating the promising coating and isolating properties. Energy-dispersive X-ray spectroscopy (EDS), X-ray photoelectron spectroscopy (XPS), and inductively coupled plasma-optical emission spectrometer (ICP-OES) showed that no metal ions escaped from the metal to the chip surface. Fourier-transform infrared spectroscopy (FTIR) was used to check the surface chemical state before and after tests, and the unchanged infrared absorption peaks indicated the unreacted, antifouling surface. The limit of detection (LOD) of at least two copies can be obtained in this chip.

23 May 2024 in ACS Biomaterials Science & Engineering
Verified
Journal Article
Closed access
QXQingyue Xian
Qingyue Xian
XLXiao Luo
Xiao Luo
JZJie Zhang
Jie Zhang
YWYu Ching Wong
Yu Ching Wong
SYSiyu Yang
Siyu Yang
WWWeijia Wen
Weijia Wen