Ventricular assist devices (VADs) are implantable turbomachines that save and improve the lives of patients with severe heart failure. In the preclinical evaluation, a VAD design must be experimentally or numerically tested regarding its pump characteristics, primarily for its pressure buildup (pressure head ) since it must provide the cardiovascular system with a sufficient blood flow rate . Those pump characteristics are determined on a test bench. Here, a glycerol-water mixture is almost exclusively used as blood-analogous fluid, which should reflect the properties (density, viscosity) of blood as close as possible. However, glycerol water has some disadvantages, such as a higher density compared to real blood and a relatively high cost. Therefore, the study aimed to analyze six different blood analogous fluids to select the most suitable one in consideration of fluid handling, costs, and, most importantly, fluid properties (material and rheological). First, all fluids were mixed to achieve reference values of blood density and viscosity from the literature. Afterwards, the pump characteristics (pressure heads and efficiencies via the VAD) were experimentally and numerically determined and compared among each other and with literature values. Of all six investigated fluids, only the aqueous–polyethylene glycol 200 (PEG 200) solution matches exactly the desired blood properties, and the pump characteristics of this fluid are in the expected range for the analyzed operation point of the VAD. Another advantage is that the cost of the mixture is 35% lower compared to glycerol water. Additionally, we demonstrate that non-Newtonian flow behavior has little effect on the pump characteristics in our VAD.
For the optimization of ventricular assist devices (VADs), flow simulations are crucial. Typically, these simulations assume single-phase flow to represent blood flow. However, blood consists of plasma and blood cells, making it a multiphase flow. Cell migration in such flows leads to a heterogeneous cell distribution, significantly impacting flow dynamics, especially in narrow gaps of less than 300 m found in VADs. In these areas, cells migrate away from the walls, forming a cell-free layer, a phenomenon not usually considered in current VAD simulations. This paper addresses this gap by introducing a viscosity model that accounts for cell migration in microchannels under VAD-relevant conditions. The model is based on local particle distributions measured in a microchannels with a blood analog fluid. We developed a local viscosity distribution for flows with particles/cells and a cell-free layer, applicable to both blood and analog fluids, with particle volume fractions of up to 5%, gap heights of 150 m, and Reynolds numbers around 100. The model was validated by comparing simulation results with experimental data of blood and blood analog fluid flow on wall shear stresses and pressure losses, showing strong agreement. This model improves the accuracy of simulations by considering local viscosity changes rather than assuming a single-phase fluid. Future developments will extend the model to physiological volume fractions up to 40%.
In the present paper, we investigate how the reductions in shear stresses and pressure losses in microfluidic gaps are directly linked to the local characteristics of cell-free layers (CFLs) at channel Reynolds numbers relevant to ventricular assist device (VAD) applications. For this, detailed studies of local particle distributions of a particulate blood analog fluid are combined with wall shear stress and pressure loss measurements in two complementary set-ups with identical flow geometry, bulk Reynolds numbers and particle Reynolds numbers. For all investigated particle volume fractions of up to , reductions in the stress and pressure loss were measured in comparison to a flow of an equivalent homogeneous fluid (without particles). We could explain this due to the formation of a CFL ranging from 10 to 20 m. Variations in the channel Reynolds number between Re = 50 and 150 did not lead to measurable changes in CFL heights or stress reductions for all investigated particle volume fractions. These measurements were used to describe the complete chain of how CFL formation leads to a stress reduction, which reduces the apparent viscosity of the suspension and results in the Fåhræus–Lindqvist effect. This chain of causes was investigated for the first time for flows with high Reynolds numbers (), representing a flow regime which can be found in the narrow gaps of a VAD.
In strongly confined flow geometries, red blood cells migrate normal to the flow direction, thereby altering the flow rheology of blood. Direct optical measurements can help to gain an improved understanding of these migration processes. In the present study, we demonstrate that astigmatism particle tracking velocimetry is a suitable 3D-particle tracking method that allows to directly measure both 3D concentration and velocity distributions of red blood cells in a flow. Red blood cells assume a non-spherical shape; therefore, the influence of their orientation on the reconstruction of the out-of-plane particle position is evaluated through a ray tracing approach of synthetic, astigmatic images. While for noise-free images, the resulting absolute out-of-plane reconstruction error $$\sigma _z$$ is small for different red blood cell orientations ( $$\sigma _z$$ = 0.98 $$\upmu \text {m}$$ ), it triples for experimentally relevant signal-to-noise ratios (SNR = 1.2). Reconstruction errors are compared to those of spherical particles. Overall, both the red blood cell orientation and the increase in signal-to-noise ratio induce similar out-of-plane reconstruction error values. Experimental analyses are also performed using both a red blood cell suspension system and a refractive index-matched suspension system of identical volume fraction ( $${1.5\,\mathrm{\%}}$$ ). Comparing results from the red blood cell suspension flow with those of the particulate suspension under identical parameters for volume fraction, particle Reynolds number, and bulk Reynolds number, a similarity in lateral migration behavior is observed under the given conditions. The results indicate that the absolute out-of-plane reconstruction error in the red blood cell suspension system ( $$\sigma _z = {4.50\,\mathrm{{\upmu \text {m}}}}$$ ) is approximately 1.5 times larger compared to the refractive index-matched system. Graphical
To better understand the actual flow fields in blood-contacting mechanical medical devices, such as ventricular assist devices (VADs), accurate in vitro and in silico studies are essential. Traditionally, these studies have treated blood as a single-phase fluid. In reality, however, blood is a multi-phase fluid consisting of plasma and suspended blood cells. Cell migration leads to heterogeneous cell distribution, which significantly impacts flow dynamics, particularly in the narrow gaps of these devices. This migration is not usually considered in in vitro analyses using blood analog fluids or in in silico simulations of VADs. This study presents an advanced viscosity modeling approach that accounts for cell migration effects under gap-relevant conditions. The model is based on local particle distribution data obtained from experiments with blood and particle-laden blood analog fluids in microchannels. It is applicable to both blood and particle-laden blood analog fluids, covering gap heights of 150 µm, Reynolds numbers in the range of 50 < Re < 150 and particle volume fractions of up to 30%. Previous works have investigated blood flows with significantly lower volume fractions of up to . The model’s accuracy was validated by comparing results with experimental data on pressure losses of blood flowing through a microchannel, demonstrating good agreement. By incorporating variations in local viscosity, this enhanced viscosity distribution model improves the accuracy of flow simulations, offering a more realistic representation of blood flow in narrow gaps compared to the single-phase assumption. Future work will extend the model to accommodate physiological particle volume fractions of up to 45%.