Background: With the growing use of dental implants, there is an urgent need to determine a prosthetic placement protocol by assessing implant stability and monitoring healing. Implant Stability Quotient (ISQ) values are produced using dental non-invasive devices through resonance frequency analysis, considered as indicators for measuring primary stability (i.e., at implant placement), monitoring biological stability (osseointegration), and prosthetic loading. A systematic and detailed comparison of ISQ measurement devices, for a given patient population, is lacking in the literature. This aspect is the subject of the present work, with the devices being two that are widely used in clinical practice (Osstell® and Osseo®100). The aim of this study was to evaluate the reliability of ISQ measurement using two standard devices most commonly used in clinical practice and to highlight any differences when comparing measurements at undefined time intervals. Methods: We enrolled 50 patients (16 males and 34 females) with a mean age of 55.4 years, who indicated dental implant placement and met the inclusion criteria. The sample was divided into two equal groups based on bone density: A (D1–D2 bone density) and B (D3–D4 bone density); each had 25 patients with 40 implants. ISQ was measured using two devices: Osstell® and Osseo 100®, at different time points (A: three and B: four follow-ups). Results: All enrolled patients completed the study without adverse events; all implants placed were successful, with no implant failure. In each of the study groups, ISQ values increased gradually with increasing follow-up time, and there was no significant difference between Osstell and Osseo 100 values at follow-up times except for the T1 follow-up in group A. Temporal comparisons for the two devices revealed significant differences in T0 vs. T2 in group A, whereas significant differences existed in T0 vs. T1, T2, and T3 in group B. Our findings indicated that the overall effect significantly depended on bone density rather than on the device used to measure ISQ. Conclusion: Regardless of the devices used, the ISQ measurement effectively monitors healing after implant insertion and allows prosthetic load to be modulated according to the ISQ value, especially when prosthetizing implants placed in fine trabecular bone (D4 or regenerated bone).
Purpose: To evaluate the effectiveness of immediate implant placement and loading combined with autologous dentin grafting in reducing post-extraction tissue remodelling, with particular emphasis on peri-implant soft tissue stability in the esthetic zone. Materials and Methods: This retrospective observational study included 13 patients (14 implant sites) with Type 1 extraction sockets in the anterior maxilla. Following atraumatic tooth extraction, implants were immediately placed and grafted with autologous dentin processed from the extracted tooth using the Tooth Transformer device. Screw-retained provisional restorations were immediately delivered through a digital workflow, and definitive restorations replicated the provisional transmucosal profile. Soft tissue dimensional changes were evaluated by superimposing stereolithography (STL) files from intraoral scans acquired before treatment and after at least 12 months. Vertical changes were measured as a single value, while horizontal changes were assessed 2 and 4 mm apical to the gingival margin. Results: All implants remained clinically stable, with no biological or prosthetic complications. Mean vertical soft tissue collapse was 0.69 mm, while mean horizontal collapse was 0.71 mm at 2 mm and 0.57 mm at 4 mm apical to the gingival margin. In most cases, dimensional changes were below 1 mm, with greater contour alterations at the more coronal level. Conclusions: Within the limitations of this study, immediate implant placement and loading combined with autologous dentin grafting using the Tooth Transformer device resulted in limited peri-implant soft tissue alterations after 12 months. This protocol may represent a biologically favourable, minimally invasive approach for ridge preservation while avoiding donor site morbidity and heterologous grafting materials.