Kilovolt x-rays are clearly suboptimal compared to MV photon beams for radiotherapy of deep-seated tumours because of the increased attenuation in tissue, causing a rapid dose fall-off. This picture could change drastically when tumours can be labelled with contrast medium, containing high atomic number elements. This causes a significant dose enhancement to the tumour by exploiting the high cross sections for the photo-electric effect for kV x-rays. In this work, we have investigated the dosimetric and microdosimetric characteristics of kV contrast-enhanced radiation therapy (CERT) for different photon energies, contrast-medium concentrations and types (I and Gd). Two idealized patient treatment plans (head and lung) for irradiation with CT-arc beams were calculated. It is shown that the dose enhancement in tumours can be highly significant (up to about sixfold for realistic 80-120 kVp x-ray spectra and an iodine concentration of 50 mg ml-1) but that dose homogeneity in the tumour depends on photon energy, contrast-medium concentration and type, and irradiation scheme. An attempt to optimize the irradiation scheme is discussed. The microdosimetric study of the dose mean lineal energy shows that radiation quality changes in the contrast-medium-labelled region compared to homogeneous tissue are fairly small and limited to 10%. It is concluded that kV-CERT is a promising radiotherapy technique, provided contrast medium can be delivered reliably to tumours.
This work compares Monte Carlo (MC) dose calculations for (125)I and (103)Pd low-dose rate (LDR) brachytherapy sources performed in virtual phantoms containing a series of human soft tissues of interest for brachytherapy. The geometries are segmented (tissue type and density assignment) based on simulated single energy computed tomography (SECT) and dual energy (DECT) images, as well as the all-water TG-43 approach. Accuracy is evaluated by comparison to a reference MC dose calculation performed in the same phantoms, where each voxel's material properties are assigned with exactly known values. The objective is to assess potential dose calculation accuracy gains from DECT. A CT imaging simulation package, ImaSim, is used to generate CT images of calibration and dose calculation phantoms at 80, 120, and 140 kVp. From the high and low energy images electron density ρ(e) and atomic number Z are obtained using a DECT algorithm. Following a correction derived from scans of the calibration phantom, accuracy on Z and ρ(e) of ±1% is obtained for all soft tissues with atomic number Z ∊ [6,8] except lung. GEANT4 MC dose calculations based on DECT segmentation agreed with the reference within ±4% for (103)Pd, the most sensitive source to tissue misassignments. SECT segmentation with three tissue bins as well as the TG-43 approach showed inferior accuracy with errors of up to 20%. Using seven tissue bins in our SECT segmentation brought errors within ±10% for (103)Pd. In general (125)I dose calculations showed higher accuracy than (103)Pd. Simulated image noise was found to decrease DECT accuracy by 3-4%. Our findings suggest that DECT-based segmentation yields improved accuracy when compared to SECT segmentation with seven tissue bins in LDR brachytherapy dose calculation for the specific case of our non-anthropomorphic phantom. The validity of our conclusions for clinical geometry as well as the importance of image noise in the tissue segmentation procedure deserves further experimental investigation.
The use of low-energy photon emitters for brachytherapy applications, as in the treatment of the prostate or of eye tumours, has drastically increased in the last few years. New seed models for 103Pd and 125I have recently been introduced. The American Association of Physicists in Medicine recommends that measurements are made to obtain the dose rate constant, the radial dose function and the anisotropy function. These results must then be compared with Monte Carlo calculations to finally obtain the dosimetric parameters in liquid water. We have used the results obtained during the characterization of the new InterSource (furnished by IBt, Seneffe, Belgium) palladium and iodine sources to compare two Monte Carlo codes against experiment for these low energies. The measurements have been performed in three different media: two solid water plastics, WT1 and RW1, and polymethylmetacrylate. The Monte Carlo calculations were made using two different codes: MCNP4C and EGSnrc. These codes use photon cross-section data of a different origin. Differences were observed between both sets of input data below 100 keV, especially for the photoelectric effect. We obtained differences in the radial dose functions calculated with each code, which can be explained by the difference between the input data. New cross-section data were then tested for both codes. The agreement between the calculations using these new libraries is excellent. The differences are within the statistical uncertainties of the calculations. These results were compared with the experimental data. A good agreement is reached for both isotopes and in the three phantoms when the measured values are corrected for the presence of the TLDs in the phantom.