A low-cost video laryngoscope (VDL) called Hybrid 1.0 was developed using smart devices for visualization. To test its performance, we compared it with a high-end VDL device, using both in vitro and in vivo studies. During the in vitro study, medical students without experience in airway intubation were randomly asked to intubate a mannequin with different degrees of difficulty (Cormack–Lehane scales) by using either the Hybrid 1.0 VDL (GI) or a conventional laryngoscope (GII). During the in vivo study, N = 60 endotracheal intubations were performed by resident and base physicians, divided into two groups; the first group intubated with the Hybrid 1.0 VDL (GI) while the second group used a VDL C-Mac shovel (GII). As performance indexes, both studies reported the number of successful intubations (correct capnography signal) and intubation time. For the in vitro testing, no statistically significant differences were found regarding the number of successful intubations, while statistically significant differences were found regarding the intubation times. During the in vivo tests, procedures were performed by residents and by base physicians, and no statistically significant differences were found. The provided results point out that the VDL proposed can be clinically useful and offers technical characteristics similar to other VDLs that currently exist on the market.

28 January 2025 in Preprints.org
Verified
Preprint
Open access
PDF file
VVVictor Maya Venegas
Victor Maya Venegas
AGAbel Garcia
Abel Garcia
PHPaul Hernández-Herrera
Paul Hernández-Herrera
SJSergio Camacho Juárez
Sergio Camacho Juárez
SMSharon Macias
Sharon Macias
BRBersain Reyes
Bersain Reyes
This project explores the development of a 1D version of the U-net convolutional neural network (CNN) for automated segmentation of heart sounds from phonocardiogram (PCG) signals. The approach begins with a feature extraction process designed to identify the onset and offset of each heart sound (HS) as binary markers. This process is similar to the image segmentation techniques used in biomedical image analysis, where 80 PCG signals, each 30 seconds long and annotated with heart sound positions, were used to train the CNN. The model receives PCG signals as input and returns the exact onset and offset for each heart sound. By combining classical feature extraction with advanced image-based segmentation techniques, this method offers a more robust and scalable solution for automated heart sound analysis. The results show that the trained model outperforms traditional segmentation methods, demonstrating the potential for deep learning architectures to be adapted for 1D biomedical signal analysis.