Purpose: This case report discusses the safe delivery of total body irradiation (TBI) to a patient with a left ventricular assist device (LVAD). This treatment required radiation-dose determinations and consequential reductions for the heart, LVAD, and an external controller connected to the LVAD.Methods: The patient was treated using a traditional 16MV anterior posterior (AP)/posterior anterior (PA) technique at a source to surface distance (SSD) of 515cm for 400cGy in two fractions. A 3cm thick Cerrobend block was placed on the spoiler to reduce dose to the heart and LVAD to 150cGy. The external controller was placed in a 1cm thick plastic box to reduce neutron dose and positioned as far away as possible, just outside the treatment fields. In vivo measurements were made using optically stimulated luminescence dosimeters (OSLDs) placed inside the box at distances of 2cm, 8.5cm, and 14cm from the field edge, and on the patient along the central axis and centered behind the LVAD block. Further ion chamber measurements were made using a solid water phantom to more accurately estimate the dose delivered to the LVAD.Results: The total estimated dose to the controller ranged from 135.3cGy to 91.5cGy. The LVAD block reduced the surface dose to the patient to 271.6cGy (68.1%). The block transmission factors of the 3cm Cerrobend block measured in the phantom were 45% at 1cm depth and decreased asymptotically to around 30% at 3cm depth. Applying these transmission factors to the in vivo measurements yielded a dose of 120cGy to the implanted device.Conclusion: Physical limitations of the controller made it impossible to completely avoid dose. Shielding is recommended. The block had limited dose reduction to the surface, due to secondary particles, but appropriately reduced the dose at 3cm and beyond. More research on LVADs dose limits is required.IntroductionThere is no cure for end stage heart failure. Currently, the only viable long-term treatment is a heart transplant. Among the absolute contraindications for a heart transplant is the presence or history of cancer (most types) within the previous 5 years due to concerns about active disease. An further complicating issue is the known cardiotoxicities of cancer therapies, including radiation, chemotherapy, and immunobiologics, all exacerbated by pre-existing cardiac morbidities1,2. Left ventricular assist devices (LVADs) help to improve survival rates in cases of heart failure and can be used as a bridge to transplant candidacy and ultimately receiving a transplant. An LVAD is an implanted mechanical pump which is directly attached to the left ventricle of the heart. The pump continuously pushes blood out of the ventricle to the aorta and the rest of the body. The LVAD consists of the mechanical pump as well as an external controller attached by a short driveline cable, 100cm in total length. Generally, the pump is powered by batteries located externally to the patient. The device can also be directly plugged into an electric outlet, if needed. As LVAD technology improves, more patients are having them implanted and survival rates with them are increasing. This leads to a growing number of patients concomitantly afflicted with cancer. To bridge patients to transplant candidacy, radiation is frequently used as part of the multidisciplinary approach to treatment.Radiation is well known to damage electronics. While the effect of radiation on other cardiac implantable electrical devices (CIED), including pacemakers and defibrillators, has been well studied, there remains limited data on radiation to LVAD3-7. Much of the existing radiation data is in the form of in-vitro studies and case reports8-15. None of these studies considered the radiation sensitivity of the LVAD external controller. Here, we report the first case of dose delivered to the controller and the first reported case of total body irradiation (TBI) to a patient with an LVAD.The nature of TBI treatments makes this an especially challenging case. The goal of the TBI is to eliminate the leukemia cells within the bone marrow, as well as suppress the immune system in order to decrease the potential for transplanted hematopoietic stem cell rejection. Traditional TBI techniques use opposing fields, AP/PA or laterals, to deliver a therapeutic dose. This does not allow for the same level of precise dose delivery as more advanced treatment techniques for localized cancers. This creates a tradeoff between limiting dose to the LVAD while ensuring sufficient dose to the surrounding bone marrow. In addition, these TBI treatments utilize large open radiation fields with sufficient flash to cover the entire patient. This is quite different from the majority of case reports in localized radiotherapy where the LVAD controller is usually far away from the radiation field. Even outside of the direct radiation field, there is out-of-field dose to consider from scatter dose and linear accelerator leakage. As a result, the external controller cannot be completely removed from all radiation. This report details the treatment approach used to manage this complex patient.Furthermore, accurately assessing the dose delivered to the LVAD during a TBI poses unique challenges. In-scatter, back-scatter, and electron contamination all contribute to uncertainty in dose calculations16,17. In addition, there is a lack of information on out-of-field doses in treatments with extended SSDs such as TBI treatments18. Therefore, direct measurements are necessary to estimate dose to the LVAD and the external controller.Methods and MaterialsPatient BackgroundThe patient in this case study is a 36-year-old male, with a history of acute myeloblastic leukemia. Chemotherapy is believed to have caused cardiomyopathy leading to cardiogenic shock with an ejection fraction of 18% and the subsequent need for a left ventricular assist device (LVAD). To bridge the gap to transplant candidacy, the patient was implanted with an Abbot...